Minnesota Administrative Rules
Agency 144 - Health Department
Chapter 4731 - RADIATION SAFETY
IRRADIATORS
Part 4731.6130 - POWER FAILURES

Universal Citation: MN Rules 4731.6130

Current through Register Vol. 49, No. 13, September 23, 2024

A. If electrical power at a panoramic irradiator is lost for longer than ten seconds, the sources must automatically return to the shielded position.

B. The lock on the door of the radiation room of a panoramic irradiator may not be deactivated by a power failure.

C. During a power failure, the area of any irradiator where sources are located may be entered only when using an operable and calibrated radiation survey meter.

Statutory Authority: MS s 144.1202; 144.1203

Disclaimer: These regulations may not be the most recent version. Minnesota may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.