Part 4675.1500 - EMERGENCY ELECTRICAL SERVICE
Universal Citation: MN Rules 4675.1500
Current through Register Vol. 49, No. 13, September 23, 2024
Subpart 1. Automatic emergency power source.
An automatic emergency power source shall be provided for essential lighting, equipment, and alarm systems which will restore the power within ten seconds in case of failure of normal power.
Subp. 2. Essential lighting.
Essential lighting includes the following:
A. exitways and exit signs;
B. surgical lights in operating rooms and
general illumination in surgical corridors;
C. laboratory, recovery room, nurses station,
and in elevator; and
D. boiler room
and near source of emergency electrical service within surgical center
building.
Subp. 3. Essential equipment and alarm systems.
Essential equipment and alarm systems:
A. nurses call and/or paging
system;
B. fire alarm system
according to State Building Code's and state fire marshal's
requirements;
C. receptacles in
operating and recovery rooms;
D.
receptacles for equipment that require continuous source of power;
and
E. elevator -- automatic or
manual transfer to emergency electrical service where elevator is used for
vertical transport of patients to operating rooms.
Statutory Authority: MS s 144.56
Disclaimer: These regulations may not be the most recent version. Minnesota may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.