ELECTRICAL SYSTEMS; NEW CONSTRUCTION
Part 4658.4625 - BEDROOM RECEPTACLES; NEW CONSTRUCTION
Universal Citation: MN Rules 4658.4625
Current through Register Vol. 49, No. 13, September 23, 2024
Each resident bedroom must have a minimum of two duplex receptacles mounted on the wall at the head of the bed, and one receptacle mounted on each side of the bed, for a total of four electrical outlets per bed. In addition to the receptacles at the head and side of each bed, each resident bedroom must be provided with two additional duplex receptacles, conveniently located on adjacent walls for resident use. At least one receptacle must be located between 16 inches and 20 inches above the floor, and be accessible for resident use.
Statutory Authority: MS s 144A.04; 144A.08
Disclaimer: These regulations may not be the most recent version. Minnesota may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.