Section 180.405 - Standard - Clinical Chemistry: Quality Control
Universal Citation: 105 MA Code of Regs 105.180
Current through Register 1531, September 27, 2024
(A) Each instrument or other device shall be recalibrated or rechecked at least once on each day of use.
(B) At least one standard and one reference sample (control) shall be included with each batch of unknown specimens where such standards and reference samples are available.
(C) Control limits for standards and reference samples shall be recorded and available for review. Records shall also include the course of action to be instituted when the results are outside the acceptable limits.
(D) Screening or qualitative chemical urinalysis shall be checked daily by use of positive and negative reference samples.
Disclaimer: These regulations may not be the most recent version. Massachusetts may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.