Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 127.000 - Licensing Of Mammography Facilities
Requirements
Section 127.019 - Equipment Quality Assurance
Universal Citation: 105 CMR 127.000 MA Code of Regs 127.019
Current through Register 1531, September 27, 2024
(A) Quality Assurance Records for Mammography X-ray Equipment.
(1) Records of all
quality assurance tests shall be maintained for mammographic x-ray equipment
and shall include at least the following:
(a)
The mammographic x-ray equipment performance evaluation and optimization,
including initial acceptance testing (to be completed before patients are
exposed with the machine) and subsequent evaluations and testing;
(b) Verification that mammographic x-ray
equipment is in safe operating order;
(c) Subsequent quality assurance test
results;
(d) Notes regarding
changes in operating conditions shall be recorded on the appropriate control
charts;
(e) Operating levels and
control limits for all quality assurance tests shall be indicated on the
appropriate control chart;
(f) All
records shall bear the date the procedure was done and the initials or name of
the individual performing the test;
(g) Quality assurance tests on the
Radiologist workstation monitors shall be done in accordance with
manufacturer's requirements;
(h)
Quality assurance tests on laser hardcopy printers shall be done in accordance
with manufacturer's requirements.
(2) Records or written logs of maintenance
and/or repairs of mammographic x-ray equipment shall be kept for at least three
years. All records for a machine shall be kept in a single file at the machine
use location. Repair records shall be referenced to quality assurance test
results, if applicable.
(3) All
quality assurance records shall be maintained for a minimum of 24 months and
readily available at the use location for review by representatives of the
Department.
(B) Quality Assurance Records and Requirements for Test Equipment and Mammography Systems.
(1) The
quality assurance test equipment shall be in good operating order.
(2) Records of each repair and/or calibration
of quality assurance test equipment shall include the date of repair and/or
calibration.
(3) Records of each
repair and/or calibration of quality assurance test equipment shall be kept for
at least three years.
(4) Quality
assurance tests shall be performed in accordance with manufacturers'
recommendations after the repair and/or replacement of any component of the
mammographic x-ray system, prior to using the equipment.
(C) Quality Assurance Tests for Equipment.
(1)
Laser
Hardcopy Printer. Quality assurance tests for laser printers, if
used by the mammography facility for clinical diagnosis, shall be performed in
accordance with the manufacturer's recommendations.
(2)
Mammography
Phantom.
(a) Quality assurance
tests for image quality in fixed mammography facilities shall be performed
weekly.
(b) Quality assurance tests
for image quality in mobile mammography facilities shall be performed each time
the machine has been moved, before beginning examinations of patient at least
weekly.
(c) Each facility shall
utilize a mammography phantom approved or accepted by ACR for performing
quality assurance tests for image quality. The following image quality
requirements shall be met, for different mammography modalities:
1. For Digital Mammography, manufacturer's
specifications or specifications stated in the 2016 ACR Digital
Mammography Quality Control Manual, or its revisions.
2. For Digital Breast Tomosynthesis,
manufacturer's specifications.
3.
For stereotactic breast biopsy, manufacturer's specifications or specifications
stated in the 1999 ACR Sterotactic Breast Biopsy Quality Control
Manual, or its revisions.
(d) Each facility shall take corrective
action in the event that quality assurance tests for image quality do not meet
the requirements in 105 CMR 127.019(C)(2)(c).
(3)
Retake Analysis.
(a) Quality assurance tests for retake
analysis shall be performed as specified by the mammography system manufacturer
or at least quarterly.
(b) The test
tools for performing quality assurance tests on retake analysis are records of
rejected images.
(c)
1. Technique;
2. Motion;
3. Positioning error;
4. Digital imaging processing; and
5. Clinically significant artifacts (dead
pixel clusters on digital detector arrays, shifts of masked spots,
etc.).
(4)
Compression in the Manual
and/or Power Modes
(a) Quality
assurance tests for compression in the manual and/or power modes in fixed
mammography facilities shall be performed semi-annually.
(b) Quality assurance tests for compression
in the manual and/or power modes in mobile mammography facilities shall be
performed each time the machine has been moved, but not less than
semi-annually.
(c) Each facility
shall use a push-pull force gauge or a flat bathroom type scale method with
accuracy of + 2% of full scale for performing quality assurance
tests.
(d) When performing the
quality assurance test for compression, the maximum compression force shall
range from a minimum of 25 to a maximum of 45 pounds in both the manual and
power drive modes and shall remain consistent throughout the period of
exposure. The compression release mechanism shall be tested at the same
interval.
(5)
Automatic Exposure Control.
(a) Quality assurance tests for automatic
exposure control (phototiming) at fixed mammography facilities shall be
performed at least annually.
(b)
Quality assurance tests for automatic exposure control (phototiming) at mobile
mammography facilities shall be performed at least semi-annually.
(6)
Kilovolt Peak
(kVp) Accuracy.
(a) Quality
assurance tests for kVp accuracy at fixed mammography facilities shall be
performed at least annually.
(b)
Quality assurance tests for kVp accuracy at mobile mammography facilities shall
be performed at least semi-annually.
(c) Each facility shall use a non-invasive
kVp meter with appropriate calibration procedures or an oscilloscope and
voltage divider network to perform quality assurance tests.
(d) When performing quality assurance tests
for kVp accuracy, the actual (measured) versus indicated kVp shall be
maintained as specified by the x-ray equipment manufacturer, but shall not vary
from the indicated kVp by greater than 5% from the indicated.
(7)
Glandular Tissue
Dose.
(a) Quality assurance tests
for glandular tissue dose at fixed mammography facilities shall be performed at
least annually.
(b) Quality
assurance tests for glandular tissue dose at mobile mammography facilities
shall be performed at least semi-annually.
(c) Each facility shall use an ion chamber or
solid state detector calibrated at the energy ranges used in mammography for
performing quality assurance tests.
(d) When performing quality assurance tests
for glandular tissue dose, the technique setting shall be those used for
imaging a 4.5 cm compressed breast in the cranio-caudal projection and the
measurement point shall be 4.5 cm from the surface of the breast support as
appropriate.
(e) For mammographic
examination of a 4.5 centimeter (cm) thick, compressed breast, consisting of
50% glandular and 50% adipose tissue, the average glandular dose shall be no
more than 3 mGy per view.
(8)
Spatial
Resolution.
(a) Quality assurance
tests for spatial resolution at fixed mammography facilities shall be performed
at least annually.
(b) Quality
assurance tests for spatial resolution at mobile mammography facilities shall
be performed at least semi-annually.
(9)
Half-value Layer
(HVL)
(a) Quality assurance tests
for half-value layer at fixed mammography facilities shall be performed at
least annually.
(b) Quality
assurance tests for half-value layer at mobile mammography facilities shall be
performed at least semi-annually.
(c) When performing quality assurance tests
for half-value layer, the HVL shall be measured with the mammography x-ray
unit's compression paddle in place over the ionization chamber or solid state
detector.
(10) All
mammography facilities shall follow the manufacturer's recommendations or the
ACR Digital Mammography QC Manual for quality control of all
components used for digital image acquisition, digital image processing and
digital image display.
Disclaimer: These regulations may not be the most recent version. Massachusetts may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.