Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 127.000 - Licensing Of Mammography Facilities
Requirements
Section 127.016 - Optimum Exposure Ranges
Universal Citation: 105 CMR 127.000 MA Code of Regs 127.016
Current through Register 1531, September 27, 2024
For mammographic examination of a 4.5 centimeter (cm) thick, compressed breast, consisting of 50% glandular and 50% adipose tissue, the average glandular dose shall be no more than 3mGy as measured using mammography phantom accepted by the U.S. Food and Drug Administration.
Disclaimer: These regulations may not be the most recent version. Massachusetts may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.