Title 105 CMR 120.000 - The Control of Radiation
REPORTS
Section 120.709 - Operating Procedures

Current through Register 1531, September 27, 2024

(A) Particle accelerators, when not in operation, shall be secured to prevent unauthorized use.

(B) The safety interlock system shall not be used to turn off the accelerator beam except in an emergency.

(C) All safety and warning devices, including interlocks, shall be checked for proper operation at intervals not to exceed three months. Results of such tests shall be maintained at the accelerator facility for inspection by the Agency.

(D) Electrical circuit diagrams of the accelerator and the associated safety interlock systems shall be kept current and maintained for inspection by the Agency and shall be available to the operator at each accelerator facility.

(E) If, for any reason, it is necessary to intentionally bypass a safety interlock or interlocks, such action shall be:

(1) Authorized by the radiation safety committee and/or radiation safety officer;

(2) Recorded in a permanent log and a notice posted at the accelerator control console; and,

(3) Terminated as soon as possible.

(F) A copy of the current operating and the emergency procedures shall be maintained at the accelerator control panel.

Disclaimer: These regulations may not be the most recent version. Massachusetts may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.