Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 120.000 - The Control of Radiation
REPORTS
Section 120.641 - Performance Criteria for Sealed Sources
Current through Register 1531, September 27, 2024
(A) Requirements for sealed sources installed after July 1, 1993:
(B) Temperature: The test source must be held at -40 E C for 20 minutes, 600 E C for one hour, and then be subjected to thermal shock test with a temperature drop from 600 E C to 20 E C within 15 seconds.
(C) Pressure: The test source must be twice subjected for at least five minutes to an external pressure (absolute) of two million newtons per square meter.
(D) Impact: A 2-kilogram steel weight, 2.5 centimeters in diameter, must be dropped from a height of one meter onto the test source.
(E) Vibration: The test source must be subjected three times for ten minutes each to vibrations sweeping from 25 hertz to 500 hertz with a peak amplitude of five times the acceleration of gravity. In addition, each test source must be vibrated for 30 minutes at each resonant frequency found.
(F) Puncture: A 50-gram weight and pin, 0.3-centimeter pin diameter, must be dropped from a height of one meter onto the test source.
(G) Bend: If the length of the source is more than 15 times larger than the minimum cross-sectional dimension, the test source must be subjected to a force of 2000 newtons at its center equidistant from two support cylinders, the distance between which is ten times the minimum cross-sectional dimension of the source.