Section 120.438 - Calibration of Survey Instruments
Current through Register 1531, September 27, 2024
(A) The registrant shall ensure that the survey instruments used to show compliance with 105 CMR 120.430 have been calibrated before first use, at intervals not to exceed 12 months, and following repair.
(B) To satisfy the requirements of 105 CMR 120.438(A), the registrant shall:
(C) To satisfy the requirements of 105 CMR 120.438(B), the registrant shall:
(D) The registrant shall retain a record of each calibration required in 105 CMR 120.438(A) for three years. The record shall include:
(E) The registrant may obtain the services of individuals licensed by the Agency, the U.S. Nuclear Regulatory Commission, an Agreement State, or a Licensing State to perform calibrations of survey instruments. Records of calibrations which contain information required by 105 CMR 120.438(D) shall be maintained by the registrant.