Current through Register 1531, September 27, 2024
(A)
Beam Limitation Except for Mammographic Systems. The
useful beam shall be limited to the area of clinical interest. This shall be
deemed to have been met if a positive beam limiting device meeting
manufacturer's specifications and the requirements of 105 CMR 120.406(G)(2) has
been properly used or if evidence of collimation is shown on at least three
sides or three corners of the film (for example, projections from the shutters
of the collimator, cone cutting at the corners, or borders at the film's edge).
(1)
General Purpose Stationary
and Mobile X-ray Systems, Including Veterinary Systems (Other than Portable)
Installed After December 31, 1997.
(a) Only x-ray systems provided with a means
for independent stepless adjustment of at least two dimensions of the x-ray
field shall be used;
(b) A method
shall be provided for visually defining the perimeter of the x-ray field. The
total misalignment of the edges of the visually defined field with the
respective edges of the x-ray field along either the length or width of the
visually defined field shall not exceed 2% of the distance from the source to
the center of the visually defined field when the surface upon which it appears
is perpendicular to the axis of the x-ray beam;
(c) The Agency may grant an exemption on
non-certified x-ray systems to 105 CMR 120.406(A)(1)(a); and,
(d) Provided the registrant makes a written
application for such exemption and in that application:
1. Demonstrates it is impractical to comply
with 105 CMR 120.406(A)(1)(a) and (b); and,
2. The purpose of 105 CMR 120.406(A)(1)(a)
and (b) will be met by other methods.
(2)
Additional Requirements for
Stationary General Purpose X-ray Systems. In addition to the
requirements of 105 CMR 120.406, all stationary general purpose x-ray systems,
both certified and non-certified shall meet the following requirements:
(a) A method shall be provided to indicate
when the axis of the x-ray beam is perpendicular to the plane of the image
receptor, to align the center of the x-ray field with respect to the center of
the image receptor to within two percent of the SID, and to indicate the SID to
within 2%;
(b) The beam-limiting
device shall indicate numerically the field size in the plane of the image
receptor to which it is adjusted; and,
(c) Indication of field size dimensions and
SID's shall be specified in inches and/or centimeters, and shall be such that
aperture adjustments result in x-ray field dimensions in the plane of the image
receptor which correspond to those indicated by the beam-limiting device to
within 2% of the SID when the beam axis is indicated to be perpendicular to the
plane of the image receptor.
(3)
X-ray Systems Designed for
One Image Receptor Size. Radiographic equipment designed for only
one image receptor size at a fixed SID shall be provided with means to limit
the field at the plane of the image receptor to dimensions no greater than
those of the image receptor, and to align the center of the x-ray field with
the center of the image receptor to within 2% of the SID, or shall be provided
with means to both size and align the x-ray field such that the x-ray field at
the plane of the image receptor does not extend beyond any edge of the image
receptor.
(4)
Radiographic Systems Other Than Those Designated in 105 CMR
120.406(A)(1) through (3).
(a)
Means shall be provided to limit the x-ray field in the plane of the image
receptor so that such field does not exceed each dimension of the image
receptor by more than 2% of the SID when the axis of the x-ray beam is
perpendicular to the plane of the image receptor.
(b) Means shall be provided to align the
center of the x-ray field with the center of the image receptor to within 2% of
the SID, or means shall be provided to both size and align the x-ray field such
that the x-ray field at the plane of the image receptor does not extend beyond
any edge of the image receptor. Compliance shall be determined with the axis of
the x-ray beam perpendicular to the plane of the image receptor.
(c)105 CMR 120.406(A)(4)(a) and (b) may be
met with a system that meets the requirements for a general purpose x-ray
system as specified in 105 CMR 120.406(A)(1) or, when alignment means are also
provided, may be met with either:
1. An
assortment of removable, fixed-aperture, beam-limiting devices sufficient to
meet the requirement for each combination of image receptor size and SID for
which the unit is designed with each such device having clear and permanent
markings to indicate the image receptor size and SID for which it is designed;
or,
2. A beam-limiting device
having multiple fixed apertures sufficient to meet the requirement for each
combination of image receptor size and SID for which the unit is designed.
Permanent, clearly legible markings shall indicate the image receptor size and
SID for which each aperture is designed and shall indicate which aperture is in
position for use.
(B)
Radiation Exposure
Control.
(1)
Timers. Means shall be provided to initiate the
radiation exposure by a deliberate action on the part of the operator, such as
the depression of a switch. Radiation exposure shall not be initiated without
such an action. In addition, it shall not be possible to initiate an exposure
when the timer is set to a "zero" or "off" position if either position is
provided.
(2)
Exposure
Indication. Means shall be provided for visual indication
observable at or from the operator's protected position whenever x-rays are
produced. In addition, a signal audible to the operator shall indicate that the
exposure has terminated.
(3)
Exposure Termination. Means shall be provided to
terminate the exposure at a preset time interval, preset product of current and
time, a preset number of pulses, or a preset radiation exposure to the image
receptor. Except for dental panoramic systems, termination of an exposure shall
cause automatic resetting of the timer to its initial setting or to "zero."
(a)
Manual Exposure
Control.
An x-ray control shall be incorporated into each x-ray system
such that an exposure can be terminated by the operator at any time except
for:
1. Exposure 1/2 of second or
less; or,
2. During serial
radiography when means shall be provided to permit completion of any single
exposure of the series in process.
(b)
Automatic Exposure
Controls. When an automatic
Exposure
control is provided:
1. Indication shall be
made on the control panel when this mode of operation is selected;
2. If the x-ray tube potential is equal to or
greater than 50 kVp, the minimum exposure time for field emission equipment
rated for pulsed operation shall be equal to or less than a time interval
equivalent to two pulses;
3. The
minimum exposure time for all equipment other than that specified in 105 CMR
120.406(B)(3)(b)2. shall be equal to or less than 1/60 second or a time
interval required to deliver five mAs, whichever is greater;
4. Either the product of peak x-ray tube
potential, current, and exposure time shall be limited to not more than 60 kWs
per exposure, or the product of x-ray tube current and exposure time shall be
limited to not more than 600 mAs per exposure except that, when the x-ray tube
potential is less than 50 kVp, the product of x-ray tube current and exposure
time shall be limited to not more than 2000 mAs per exposure; and,
5. A visible signal shall indicate when an
exposure has been terminated at the limits required by 105 CMR
120.406(B)(3)(b)4., and manual resetting shall be required before further
automatically timed exposures can be made.
(4)
Exposure Duration (Timer)
Linearity. For systems having independent selection of exposure
time settings, the average ratios (Xi) of exposure to
the indicated timer setting, in units of C kg
-1s
-1 (mR/s), obtained at any two
clinically used timer settings shall not differ by more than 0.10 times their
sum. This is written as:
(X1 - X2) <=
0.1 (X1 + X2)
where X1 and X2
are the average C kg-1s-1
(mR/s) values.
(5)
Exposure Control Location. The x-ray exposure control
shall be so placed that the operator can view the patient while making any
exposure.
(6)
Operator
Protection, Except Veterinary Systems.
(a)
Stationary
Systems. Stationary x-ray systems shall be required to have the
x-ray control permanently mounted in a protected area so that the operator is
required to remain in that protected area during the entire exposure.
(b)
Mobile and Portable
Systems. Mobile and portable x-ray systems which are:
1. Used continuously for greater than one
week in the same location, i.e., a room or suite, shall meet
the requirements of 105 CMR 120.406(B)(6)(a);
2. Used for less than one week at the same
location shall be provided with either a protective barrier at least 6.5 feet
(two m) high for operator protection during exposure, or means shall be
provided to allow the operator to be at least nine feet (2.7 m) from the tube
housing assembly during the exposure.
3. The x-ray control shall provide visual
indication observable at or from the operator's protected position whenever
x-rays are produced. In addition, a signal audible to the operator shall
indicate that the exposure has terminated.
(7)
Exposure
Reproducibility. When all techniques factors are held constant,
including control panel selections associated with exposure control systems,
the coefficient of variation of exposure for both manual and automatic exposure
control systems shall not exceed 0.05. This requirement applies to clinically
used techniques.
(C)
Source-to-skin Distance. All mobile or portable
radiographic systems shall be provided with means to limit the source-to-skin
distance to equal to or greater than 30 centimeters, except for veterinary
systems.
(D)
Radiation
from Capacitor Energy Storage Equipment in Standby Status.
Radiation emitted from the x-ray tube when the exposure switch or timer is not
activated shall not exceed a rate of two milliroentgen (0.516 µC/kg) per
hour at five centimeters from any accessible surface of the diagnostic source
assembly, with the beam-limiting device fully open.
(E)
Accuracy.
Deviation of measured technique factors from indicated values of kVp and
exposure time shall not exceed the limits specified for that system by its
manufacturer. In the absence of manufacturer's specifications, the deviation
shall not exceed 10% of the indicated value for kVp and 20% for time.
(F)
mA/mAs
Linearity. The following requirements apply when the equipment is
operated on a power supply as specified by the manufacturer for any fixed x-ray
tube potential within the range of 40% to 100% of the maximum rated:
(1)
Equipment Having Independent
Selection of X-ray Tube Current (mA). The average ratios
(Xi) of air kerma to the indicated milliampere-seconds
product (C kg-1 mAs-1 (or
mR/mAs)) obtained at any two consecutive tube current settings shall not differ
by more than 0.10 times their sum:
X1-X2 <=
0.10 (X1+X2)
where X1 and X2
are the average values obtained at each of two consecutive tube current
settings, or at two settings differing by no more than a factor of two where
the tube current selection is continuous.
(2)
Equipment Having a Combined
X-ray Tube Current-exposure Time Product (mAs) Selector. The
average ratios (Xi) of air kerma to the indicated
milliampere-seconds product, in units of C kg-1
mAs-1 (or mR/mAs), obtained at any two consecutive
mAs selector settings shall not differ by more than 0.10 times their sum:
X1-X2 <=
0.10 (X1 +X2)
where X1 and X2
are the average values obtained at any two mAs selector settings, or at two
settings differing by no more than a factor of two where the mAs selector
provides continuous selection.
(3)
Measuring
Compliance. Determination of compliance shall be based on three
exposures taken within a time period of one hour, at each of the two settings.
These two settings may include any two focal spot sizes except where one is
equal to or less than 0.45 millimeters and the other is greater than 0.45
millimeters. For purposes of this requirement, focal spot size is the nominal
focal spot size specified by the x-ray tube manufacturer.
(G)
Additional Requirements
Applicable to Certified Systems Only. Diagnostic x-ray systems
incorporating one or more certified component(s) shall be required to comply
with the following additional requirement(s) which relate to that certified
component(s).
(1)
Beam Limitation
for Portable X-ray Systems. Beam limitation for portable x-ray
systems shall meet the beam limitation requirements of 105 CMR
120.406(A)(1).
(2)
Field Limitation and Alignment on Stationary General Purpose X-ray
Systems. For stationary, general purpose x-ray systems which
contain a tube housing assembly, an x-ray control, and, for those systems so
equipped, a table, all certified in accordance with
21 CFR
1020.30(C):
(a) Means shall be provided for positive beam
limitation which will, at the SID for which the device is designed, either
cause automatic adjustment of the x-ray field in the plane of the image
receptor to the image receptor size within five seconds after insertion of the
image receptor or, if adjustment is accomplished automatically in a time
interval greater than five seconds or is manual, will prevent production of
x-rays until such adjustment is completed. For the SID at which the device is
not intended to operate, the device shall prevent the production of
x-rays.
(b) The x-ray field size in
the plane of the image receptor, whether automatically or manually adjusted,
shall be such that neither the length nor the width of the x-ray field differs
from that of the image receptor by greater than 2% of the SID and that the sum
of the length and width differences without regard to sign be no greater than
3% of the SID when the equipment indicates that the beam axis is perpendicular
to the plane of the image receptor.
(c) The radiographic system shall be capable
of operation, at the discretion of the operator, such that the field size at
the image receptor can be adjusted to a size smaller than the image receptor.
The minimum field size at a distance of 100 centimeters shall be equal to or
less than five by five centimeters. Return to positive beam limitation as
specified in 105 CMR 120.406(F)(6)(a) and (b) shall occur upon a change in
image receptor.
(d) Positive beam
limitation may be bypassed when radiography is conducted which does not use the
cassette tray or permanently mounted vertical cassette holder, or when either
the beam axis or table angulation is not within 10° of the horizontal or
vertical during any part of the exposure, or during stereoscopic radiography.
If the bypass mode is provided, return to positive beam limitation shall be
automatic.
(e) A capability may be
provided for overriding positive beam limitation in the event of system failure
or to perform special procedures which cannot be performed in the positive
mode. If so provided, a key shall be required to override the positive mode It
shall be impossible to remove the key while the positive mode is
overridden.
(3)
Timers. Except for dental panoramic systems,
termination of exposure shall cause automatic resetting of the timer to its
initial setting or to "zero."
(4)
Transmission Limit for Image Receptor Supporting Devices Used for
Mammography. For x-ray systems manufactured after September 5,
1978 which are designed only for mammography, the transmission of the primary
beam through any image receptor support provided with the system shall be
limited such that the Exposure five centimeters from
any accessible surface beyond the plane of the image receptor supporting device
does not exceed 0.1 milliroentgen (25.8 C/kg) for each activation of the tube.
Exposure shall be measured with the system operated at
the minimum SID for which it is designed. Compliance shall be determined at the
maximum rated peak tube potential for the system and at the maximum rated
product of tube current and exposure time (mAs) for that peak tube potential.
Compliance shall be determined by measurements averaged over an area of 100
square centimeters with no linear dimension greater than 20
centimeters.
(H) Any
facility that utilizes a CR or DR system shall follow the manufacturer's
recommendations for Quality Assurance and Quality Control.
(1) All Quality Control tests and results
shall be documented.
(2) Quality
assurance and quality control records shall be maintained for a minimum of 24
months and readily available for review by representatives of the
Department.