(1)
Registrant. The
registrant shall be responsible for directing the operation of the x-ray
system(s) under his administrative control. The registrant or the registrant's
agent shall assure that the requirements of 105 CMR 120.403(A)(1) are met in
the operation of the x-ray system(s).
(a) An
x-ray system which does not meet the provisions of 105 CMR 120.400 shall not be
operated for diagnostic or therapeutic purposes, unless otherwise approved by
the Radiation Control Program.
(b)
Individuals who shall be operating the x-ray systems shall meet the
requirements of
105 CMR 125.000:
Licensing of Radiologic Technologists.
(c) A chart shall be provided in the vicinity
of the diagnostic x-ray system's control panel which specifies, for all
examinations performed with that system, the following information:
1. Patient's body size and anatomical size,
or body part thickness, or age (for pediatrics), versus technique factors to be
utilized;
2. Type and size of the
image receptor to be used;
3. Type
and size of the image receptor combination to be used; if any
4. Source to image receptor distance to be
used (except for dental intra-oral radiography); and
5. Type and location of placement of patient
shielding (e.g. gonad, etc.) to be used.
(d) The registrant of a facility shall create
and make available to x-ray operators written safety procedures, including
patient holding and any restrictions of the operating technique required for
the safe operation of the particular x-ray system. The operator shall be able
to demonstrate familiarity with these procedures. These procedures shall be
reviewed, updated, and documented annually by management.
(e) Except for patients who cannot be moved
out of the room, only the staff, ancillary personnel, and parents of pediatric
patients whose presence might be required for the medical procedure or training
shall be in the room or area during the radiographic exposure. Other than the
patient being examined:
1. All individuals
shall be positioned such that no part of the body will be struck by the useful
beam unless protected by 0.5 millimeter lead equivalent.
2. The x-ray operator, other staff, ancillary
personnel, and other persons required for the medical procedure shall be
protected from the direct scatter radiation by protective aprons or whole body
protective barriers of not less than 0.25 millimeter lead equivalent. For
interventional procedures, lead glasses shall be used.
3. Human patients who cannot be removed from
the room shall be protected from the direct scatter radiation by whole body
protective barriers of 0.25 millimeter lead equivalent or shall be so
positioned that the nearest portion of the body is at least two meters from
both the tube head and the nearest edge of the image receptor.
(f) Gonad shielding of not less
than 0.50 millimeter lead equivalent shall be used for patients, who have not
passed the reproductive age, during radiographic procedures in which the gonads
are in the useful beam, except for cases in which this would interfere with the
diagnostic procedure.
(g)
Individuals shall not be exposed to the useful beam except for healing arts
purposes and unless such exposure has been authorized by a licensed
practitioner of the healing arts. This provision specifically prohibits
deliberate exposure for the following purposes:
1. Exposure of an individual for training,
demonstration, or other non-healing-arts purposes; and
2. exposure of an individual for the purpose
of healing arts screening except as authorized by 105 CMR
120.403(A)(1)(k).
(h)
When a patient or image receptor must be provided with auxiliary support during
a radiation exposure:
1. Mechanical holding
devices shall be used when the technique permits. The written safety
procedures, required by 105 CMR 120.403(A)(1)(d), shall list individual
projections, specific patient conditions, or psychological development level
where holding devices cannot be utilized;
2. Written safety procedures, as required by
105 CMR 120.403(A)(1)(d), shall indicate the requirements for selecting a
holder and the procedure the holder shall follow;
3. The human holder shall be instructed in
personal radiation safety and protected as required by 105 CMR
120.403(A)(1)(e);
4. No individual
shall be used routinely to hold film or patients;
5. In those cases where the patient must hold
the film, except during intraoral examinations, any portion of the body other
than the area of clinical interest struck by the useful beam shall be protected
by not less than 0.5 millimeter lead equivalent material; and,
6. Each facility shall have leaded aprons and
gloves available in sufficient numbers to provide protection to all personnel
who are involved with x-ray operations and who are otherwise not
shielded.
(i) Procedures
and auxiliary equipment designed to minimize patient and personnel exposure
commensurate with the needed diagnostic information shall be utilized.
1. An imaging system of appropriate speed
consistent with the diagnostic objective of the examinations shall be used.
Film cassettes without intensifying screens shall not be used for any routine
diagnostic radiological imaging, with the exception of veterinary radiography
and standard film packets for intra-oral use in dental radiography.
2. The radiation exposure to the patient
shall be the minimum exposure required to produce images of good diagnostic
quality.
3. Protective equipment
including aprons, gloves, and shields shall be x-rayed annually for defects,
such as holes, cracks, and tears to assure reliability and integrity. A record
of this test shall be maintained for inspection by the Radiation Control
Program. If such defect is found, equipment shall be replaced or removed from
service until repaired or replaced.
4. Radiographic systems other than
fluoroscopic, dental intra-oral, or veterinarian systems shall not be utilized
in procedures where the source to patient distance is less than 30
centimeters.
5. Mammographic
procedures shall only be performed on special purpose mammographic
equipment.
6. Mobile or portable
radiographic systems shall only be used for examinations where it is
impractical to transfer the patient(s) to a stationary x-ray
installation.
7. If grids are used
between the patient and the image receptor to decrease scatter to the film and
improve contrast, the grid shall:
a. Be
positioned properly, i.e., tube facing the right direction, and grid centered
to the central ray;
b. If the grid
is of the focused type, be of the proper focal distance for the SIDs being
used.
(j) All
occupationally exposed individuals are subject to the requirements of
105 CMR
120.211, 120.215,
120.217
and
120.218.
(k) If the facility ceases to operate, the
Registrant or Responsible person of the facility must notify the Radiation
Control Program within 15 days. Included in this notification, is the name and
address of the person who disposed of the x-ray unit.
(2) The registrant of the facility shall
ensure that the equipment is in safe operating condition:
(a) when it is first installed and prior to
use on patients;
(b) after any
major changes or replacement of parts and prior to use on patients:
(c) by having physics surveys, calibrations
and preventative maintenance such physics surveys and preventative maintenance
shall be made annually:
1. The physics surveys
shall be performed by a qualified medical physicist;
2. The preventative maintenance or
calibration shall be performed by a registered service provider as specified in
105 CMR
120.026.
(d) Physics surveys shall be reviewed and
signed within a reasonable time of completion of the tests but no longer than
30 days of completion of the tests by a qualified medical physicist and a
responsible person at the facility or responsible physician, and any necessary
corrective action shall be implemented within 30 days.
(e) Records of calibrations and preventative
maintenance shall be maintained at the facility for three years.
(3)
Information and
Maintenance Record and Associated Information. The registrant of a
facility shall maintain the following information for each x-ray system for
inspection by the Radiation Control Program:
(a) Model and serial numbers of all major
components, and user's manuals for those components;
(b) Records of installation, surveys,
calibrations, maintenance, and modifications performed on the x-ray system(s)
after the effective date of 105 CMR 120.000 with the names of persons who
performed such services;
(c) A copy
of the service providers certificate of registration shall be maintained by the
facility.
(d) A scale drawing
provided by a registered service provider or qualified medical physicist of the
room in which a stationary x-ray system is located with such drawing indicating
the current use of areas adjacent to the room and an estimation of the extent
of occupancy by an individual in such areas. In addition, the drawing shall
include:
1. The results of a survey for
radiation levels present at the operator's position and at pertinent points
outside the room at specified test conditions; or
2. The type and thickness of materials, or
lead equivalency, of each protective barrier; and,
(e) A copy of all correspondence with this
Radiation Control Program regarding that x-ray system.
(6)
Quality Assurance
Program.
(a) All registrants of
diagnostic x-ray imaging equipment shall establish and maintain a quality
assurance program consisting of quality control assessments addressing at least
the following items:
1. Written standard
operating procedures on radiation protection and the practice of radiologic
technology reviewed, updated, and documented annually by management.
2. Employee review and written
acknowledgement of standard operating procedures and policies on radiation
protection and the practice of radiologic technology shall be documented
annually.
3. Credentialling of
practitioners, medical physicists, and x-ray equipment operators.
4. Film Processing equipment:
a. Compliance with 105 CMR
120.403(C);
b. Film processor
performance to include medium density, density difference, and base +
fog;
c. Darkroom fog;
5. Radiographic equipment:
a. Compliance with performance standards in
105 CMR
120.404 and
120.406;
b. Entrance skin exposure rates of selected
patient examinations;
c. Image
printing and viewing equipment;
d.
Measurement of low and high contrast resolution; and
e. Radiation protection.
6. Fluoroscopic equipment:
a. Compliance with performance standards in
105 CMR
120.405;
b. Entrance skin exposure rates of selected
patient examinations;
c. Image
printing and viewing equipment;
d.
Measurement of low and high contrast resolution; and
e. Radiation protection.
7. Computerized tomography equipment:
a. Compliance with performance standards in
105 CMR
120.409;
b. CT number;
c. Low contrast and high contrast
resolution;
d. Dosimetry of
selected patient examinations to include pediatric patients if
applicable;
e. Image printing and
viewing equipment; and
f. Radiation
protection.
8. Bone
densitometry equipment: Compliance with requirements in
105 CMR
120.410.
9. Structural shielding for new facilities
with x-ray equipment:
a. Pre-construction
shielding design and evaluation; and
b. Post-construction radiation protection
survey.
10. Structural
shielding for modifying use or equipment in existing facility:
a. Re-evaluation of shielding design;
and
b. Post-modification radiation
protection survey.
(b) The registrant of a facility shall assign
qualified personnel to fully implement the quality assurance program.
(c) Quality control assessments may be
assigned to qualified personnel who possess the requisite training and/or
experience.
(d) Quality control
assessments shall be conducted by or under the direction of, a qualified
medical physicist.
(e) The
registrant of a facility and/or qualified medical physicist shall determine the
frequency of quality control tests but shall not be less stringent than the
manufacturers recommendations.
(f)
The quality assurance program shall be in written form and available for review
by the Agency.
(g) Equipment used
for compliance with the provisions of 105 CMR 120.403(6) shall be properly
calibrated and maintained in accordance with accepted professional standards.
105 CMR 120.403(A)(6) does not pertain to quality assurance
for mammography equipment see105 CMR 127.000.