Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 120.000 - The Control of Radiation
PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 120.583 - Radiation Surveys
Current through Register 1531, September 27, 2024
(A) In addition to the survey requirements in 105 CMR 120.225, a person licensed pursuant to 105 CMR 120.500 shall make surveys to ensure that the maximum radiation levels and average radiation levels from the surface of the main source safe with the source(s) in the shielded position does not exceed the levels stated in the Sealed Source and Device Registry.
(B) The licensee shall make the survey required by 105 CMR 120.583(A) at installation of a new source and following repairs to the source(s) shielding, the source(s) driving unit, or other electronic or mechanical component that could expose the source, reduce the shielding around the source(s), or compromise the radiation safety of the unit or the source(s).
(C) A licensee shall retain a record of the radiation surveys required in 105 CMR 120.583(A) in accordance with 105 CMR 120.592(K).