Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 120.000 - The Control of Radiation
PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 120.582 - Additional Technical Requirements for Mobile Remote Afterloader Units
Current through Register 1531, September 27, 2024
(A) A licensee providing mobile remote afterloader service shall:
(B) In addition to the periodic spot-checks required by 105 CMR 120.580, a licensee authorized to use mobile afterloaders for medical use shall perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of:
(C) In addition to the requirements for checks in 105 CMR 120.582(B), a licensee shall ensure overall proper operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.
(D) If the results of the checks required in 105 CMR 120.582(B) indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
(E) A licensee shall retain a record of each check required by 105 CMR 120.582(B) in accordance with 105 CMR 120.592(J).