Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 120.000 - The Control of Radiation
PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 120.581 - Periodic Spot-checks for Gamma Stereotactic Radiosurgery Units
Current through Register 1531, September 27, 2024
(A) A licensee authorized to use a gamma stereotactic radiosurgery unit for medical use shall perform spot-checks on each gamma stereotactic radiosurgery facility and on each unit:
(B) A licensee shall have the authorized medical physicist:
(C) To satisfy the requirement of 105 CMR 120.581(A)(1), spot-checks must, at a minimum:
(D) To satisfy the requirements of 105 CMR 120.581(A)(2) and (3), spot-checks must assure proper operation of:
(E) A licensee shall arrange for prompt repair of any system identified in 105 CMR 120.581(C) that is not operating properly.
(F) If the results of the checks required in 105 CMR 120.581(D) indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
(G) A licensee shall retain a record of each check required by 105 CMR 120.581(C) and (D) in accordance with 105 CMR 120.592(I).