Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 120.000 - The Control of Radiation
PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 120.580 - Periodic Spot-checks for Remote Afterloader Units

Current through Register 1531, September 27, 2024

(A) A licensee authorized to use remote afterloader units for medical use shall perform spot-checks on each remote afterloader facility and on each unit:

(1) At the beginning of each day of use of a high dose-rate, medium dose-rate or pulsed dose-rate remote afterloader unit;

(2) Prior to each patient treatment with a low dose-rate remote afterloader unit; and

(3) After each source installation.

(B) A licensee shall have the authorized medical physicist establish written procedures for performing the spot-checks required in 105 CMR 120.580(A). The authorized medical physicist need not actually perform the spot-check measurements.

(C) A licensee shall have the authorized medical physicist review the results of each output spot-check within 15 days. The authorized medical physicist shall promptly notify the licensee in writing of the results of each output spot-check.

(D) To satisfy the requirement of 105 CMR 120.580(A), spot-checks must, at a minimum, assure proper operation of:

(1) Electrical interlocks at each remote afterloader unit room entrance;

(2) Source exposure indicator lights on the remote afterloader unit, on the control console, and in the facility;

(3) Viewing and intercom systems in each high dose-rate, medium dose-rate and pulsed dose-rate remote afterloader facility;

(4) Emergency response equipment;

(5) Radiation monitors used to indicate the source position;

(6) Timer accuracy;

(7) Clock (date and time) in the unit's computer; and

(8) Decayed source(s) activity in the unit's computer.

(E) If the results of the checks required in 105 CMR 120.580(D) indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.

(F) A licensee shall retain a record of each spot-check required by 105 CMR 120.580(D), in accordance with 105 CMR 120.592(H).

Disclaimer: These regulations may not be the most recent version. Massachusetts may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.