Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 120.000 - The Control of Radiation
PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 120.579 - Periodic Spot-checks for Teletherapy Units
Current through Register 1531, September 27, 2024
(A) A licensee authorized to use teletherapy units for medical use shall perform output spot-checks on each teletherapy unit once in each calendar month that include determination of:
(B) A licensee shall perform measurements required by 105 CMR 120.579(A) in accordance with procedures established by the authorized medical physicist. That individual need not actually perform the output spot-check measurements.
(C) A licensee shall have the authorized medical physicist review the results of each output spot-check within 15 days. The authorized medical physicist shall promptly notify the licensee in writing of the results of each output spot-check.
(D) A licensee authorized to use a teletherapy unit for medical use shall perform safety spot-checks of each teletherapy facility once in each calendar month and after each source installation to assure proper operation of:
(E) If the results of the checks required in 105 CMR 120.579(D) indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
(F) A licensee shall retain a record of each spot-check required by 105 CMR 120.579(A) and (D), in accordance with 105 CMR 120.592(G).