Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 120.000 - The Control of Radiation
PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 120.578 - Full Calibration Measurements on Gamma Stereotactic Radiosurgery Units
Current through Register 1531, September 27, 2024
(A) A licensee authorized to use a gamma stereotactic radiosurgery unit for medical use shall perform full calibration measurements on each unit:
(B) To satisfy the requirement of 105 CMR 120.578(A), full calibration measurements shall include determination of:
(C) A licensee shall use the dosimetry system described in 105 CMR 120.575(A) to measure the output for one set of exposure conditions. The remaining radiation measurements required in 105 CMR 120.578(B)(1) may be made using a dosimetry system that indicates relative dose rates.
(D) A licensee shall make full calibration measurements required by 105 CMR 120.578(A) in accordance with published protocols accepted by nationally recognized bodies.
(E) A licensee shall mathematically correct the outputs determined in 105 CMR 120.578(B)(1) at intervals not exceeding one month for cobalt-60 and at intervals consistent with 1% physical decay for all other radionuclides.
(F) Full calibration measurements required by 105 CMR 120.578(A) and physical decay corrections required by 105 CMR 120.578(E) must be performed by the authorized medical physicist.
(G) A licensee shall retain a record of each calibration in accordance with 105 CMR 120.592(F).