Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 120.000 - The Control of Radiation
PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 120.577 - Full Calibration Measurements on Remote Afterloader Units

Current through Register 1531, September 27, 2024

(A) A licensee authorized to use a remote afterloader unit for medical use shall perform full calibration measurements on each unit:

(1) Before the first medical use of the unit; and

(2) Before medical use under the following conditions:
(a) Following replacement of the source or following reinstallation of the unit in a new location outside the facility; and

(b) Following any repair of the unit that includes removal of the source or major repair of the components associated with the source exposure assembly; and

(3) At intervals not exceeding one quarter for high dose-rate, medium dose-rate, and pulsed dose-rate remote afterloader units with sources whose half-life exceeds 75 days; and

(4) At intervals not exceeding one year for low dose-rate remote afterloader units.

(B) To satisfy the requirement of 105 CMR 120.577(A), full calibration measurements shall include, as applicable, determination of:

(1) the output within +/- 5%;

(2) Source position accuracy to within +/- 1 millimeter;

(3) Source retraction with backup battery upon power failure;

(4) Length of the source transfer tubes;

(5) Timer accuracy and linearity over the typical range of use;

(6) Length of applicators; and

(7) Function of the source transfer tubes, applicators, and transfer tube-applicator interfaces.

(C) In addition to the requirements for full calibrations for low dose-rate remote afterloader units in 105 CMR 120.577(B), a licensee shall perform an autoradiograph of the source(s) to verify inventory and source(s) arrangement at intervals not exceeding one quarter.

(D) A licensee shall use the dosimetry system described in 105 CMR 120.575(A) to measure the output.

(E) A licensee shall make full calibration measurements required by 105 CMR 120.577(A) in accordance with published protocols accepted by nationally recognized bodies.

(F) For low dose-rate remote afterloader units, a licensee may use measurements provided by the source manufacturer that are made in accordance with 105 CMR 120.577(A) through (E).

(G) A licensee shall mathematically correct the outputs determined in 105 CMR 120.577(B)(1) for physical decay at intervals consistent with 1% physical decay.

(H) Full calibration measurements required by 105 CMR 120.577(A) and physical decay corrections required by 105 CMR 120.577(G) must be performed by the authorized medical physicist.

(I) A licensee shall retain a record of each calibration in accordance with 105 CMR 120.592(F).

Disclaimer: These regulations may not be the most recent version. Massachusetts may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.