Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 120.000 - The Control of Radiation
PHOTON EMITTING REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS, AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 120.577 - Full Calibration Measurements on Remote Afterloader Units
Current through Register 1531, September 27, 2024
(A) A licensee authorized to use a remote afterloader unit for medical use shall perform full calibration measurements on each unit:
(B) To satisfy the requirement of 105 CMR 120.577(A), full calibration measurements shall include, as applicable, determination of:
(C) In addition to the requirements for full calibrations for low dose-rate remote afterloader units in 105 CMR 120.577(B), a licensee shall perform an autoradiograph of the source(s) to verify inventory and source(s) arrangement at intervals not exceeding one quarter.
(D) A licensee shall use the dosimetry system described in 105 CMR 120.575(A) to measure the output.
(E) A licensee shall make full calibration measurements required by 105 CMR 120.577(A) in accordance with published protocols accepted by nationally recognized bodies.
(F) For low dose-rate remote afterloader units, a licensee may use measurements provided by the source manufacturer that are made in accordance with 105 CMR 120.577(A) through (E).
(G) A licensee shall mathematically correct the outputs determined in 105 CMR 120.577(B)(1) for physical decay at intervals consistent with 1% physical decay.
(H) Full calibration measurements required by 105 CMR 120.577(A) and physical decay corrections required by 105 CMR 120.577(G) must be performed by the authorized medical physicist.
(I) A licensee shall retain a record of each calibration in accordance with 105 CMR 120.592(F).