Code of Massachusetts Regulations
105 CMR - DEPARTMENT OF PUBLIC HEALTH
Title 105 CMR 120.000 - The Control of Radiation
GENERAL TECHNICAL REQUIREMENTS
Section 120.536 - Requirements for Possession of Sealed Sources and Brachytherapy Sources
Current through Register 1531, September 27, 2024
(A) A licensee in possession of any sealed source or brachytherapy source shall follow the radiation safety and handling instructions supplied by the manufacturer or equivalent instructions approved by the Agency.
(B) A licensee in possession of a sealed source shall:
(C) To satisfy the leak test requirements of 105 CMR 120.536(B), the licensee shall measure the sample so that the leak test can detect the presence of 185 becquerels (0.005 µCi) of radioactive material in the sample. If the leak test reveals the presence of 185 becquerels (0.005 µCi) or more of removable contamination, the licensee shall:
(D) A licensee shall retain leak test records in accordance with 105 CMR 120.590(I)(1).
(E) A licensee in possession of a sealed source or brachytherapy source, except for gamma stereotactic radiosurgery sources, shall conduct a semi-annual physical inventory of all such sources. The licensee shall retain each inventory record in accordance with 105 CMR 120.590(I)(2).