Section 120.532 - Possession, Use, and Calibration of Instruments Used to Measure the Activity of Unsealed Radioactive Material

Current through Register 1531, September 27, 2024

(A) For direct measurements performed in accordance with 105 CMR 120.534, a licensee shall possess and use instrumentation to measure the activity of unsealed radioactive materials prior to administration to each patient or human research subject.

(B) A licensee shall calibrate the instrumentation required in 105 CMR 120.532(A) in accordance with nationally recognized standards or the manufacturer's instructions.

(C) A licensee shall retain a record of each instrument calibration required by 105 CMR 120.532 in accordance with 105 CMR 120.590(F).

Disclaimer: These regulations may not be the most recent version. Massachusetts may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.