Code of Maryland Regulations
Title 10 - MARYLAND DEPARTMENT OF HEALTH
Part 3
Subtitle 10 - LABORATORIES
Chapter 10.10.13 - Medical Laboratories - Testing for Hereditary and Congenital Disorders in Newborn Infants
Section 10.10.13.20 - Quality Control
Universal Citation: MD Code Reg 10.10.13.20
Current through Register Vol. 51, No. 19, September 20, 2024
A. Quality Control Program. A permittee shall develop, establish, implement, maintain, and follow written quality control policies, procedures, and techniques that provide and ensure accurate, reliable, and valid test results and reports.
(1) A permittee shall develop, establish,
implement, maintain and follow written quality control policies, procedures,
and techniques that provide and ensure accurate, reliable, and valid test
results and reports.
(2) A
permittee's quality control program shall:
(a) Cover all tests, methods, and procedures
performed by the laboratory; and
(b) Meet the standards and requirements set
forth in this chapter and COMAR 10.10.06.
(3) A permittee's quality control procedures
and practices shall include, but are not limited to, the use of:
(a) Analyses of reference materials to ensure
the reproducibility and accuracy of test results;
(b) Systems to ensure proper functioning of
instruments, equipment, reagents, and performance of laboratory
personnel;
(c) Valid reagents,
equipment, calibration, and systematic procedural checks;
(d) A system that ensures timely and
pertinent error detection and correction for the review and reporting of test
results; and
(e) Quality control
materials from the Centers for Disease Control and Prevention's Newborn
Screening Quality Assurance Program.
B. Adequacy of Methods and Equipment. To ensure the adequacy of test methods, test systems, and equipment, a permittee shall:
(1) Employ methods, test systems, and
equipment that provide accurate and reliable test results and
reports;
(2) Have sufficient and
suitable equipment and instruments to perform the type and volume of tests
offered and performed by the laboratory;
(3) Perform tests in a manner that provides
accurate and reliable test results;
(4) Monitor test methods, instruments, and
testing systems, where applicable, for the following performance
characteristics:
(a) Method, instrument, and
test system detection limits and cutoff levels;
(b) Precision;
(c) Accuracy;
(d) Specificity;
(e) Sensitivity;
(f) Interferences; and
(g) Other related test variables that affect
test results; and
(5)
Have adequate quality control systems and procedures in place to:
(a) Obtain and verify a statistically valid
number of counts or measurement levels to give accurate and reliable test
results for an analytical instrument, each day an instrument is used for
testing; and
(b) Report results:
(i) As set forth in COMAR
10.10.06.04B;
(ii) As set forth in
42 CFR § 493.1291; and
(iii) Within the time frames required by this
chapter.
C. Supplies, Reagents, and Materials.
(1) A permittee shall ensure that supplies,
reagents, and other materials used in the collected, handling, testing, and
storing of newborn blood-spot specimens are stored, labeled, checked for proper
reactivity, and used according to the provisions of this chapter.
(2) A permittee shall:
(a) Label laboratory supplies and materials
used in the testing process, including reagents, solutions, controls, and
calibrators to indicate the:
(i) Identity of
reagents, solutions, controls, and calibrators;
(ii) Strength or concentration;
(iii) Storage requirements;
(iv) Preparation and expiration
date;
(v) Identity of the preparer;
and
(vi) Other pertinent
information that affects test performance;
(b) Store laboratory materials in the manner
recommended by the manufacturer;
(c) Test representative samples of each lot
of materials used for or in the process of testing blood-spot specimens on a
regularly scheduled basis to determine that each lot of test material is
capable of performing as required; and
(d) Use supplies, reagents, kits, test
systems, and other materials in a manner consistent with the recommendations
and instructions provided by the manufacturer.
(3) A permittee may not use supplies,
reagents, kits, test systems, or other materials in the testing process that
have:
(a) Exceeded the materials' established
expiration dates;
(b) Substandard
reactivity; or
(c)
Deteriorated.
Disclaimer: These regulations may not be the most recent version. Maryland may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.