Code of Maryland Regulations
Title 10 - MARYLAND DEPARTMENT OF HEALTH
Part 3
Subtitle 10 - LABORATORIES
Chapter 10.10.13 - Medical Laboratories - Testing for Hereditary and Congenital Disorders in Newborn Infants
Section 10.10.13.20 - Quality Control

Universal Citation: MD Code Reg 10.10.13.20

Current through Register Vol. 51, No. 19, September 20, 2024

A. Quality Control Program. A permittee shall develop, establish, implement, maintain, and follow written quality control policies, procedures, and techniques that provide and ensure accurate, reliable, and valid test results and reports.

(1) A permittee shall develop, establish, implement, maintain and follow written quality control policies, procedures, and techniques that provide and ensure accurate, reliable, and valid test results and reports.

(2) A permittee's quality control program shall:
(a) Cover all tests, methods, and procedures performed by the laboratory; and

(b) Meet the standards and requirements set forth in this chapter and COMAR 10.10.06.

(3) A permittee's quality control procedures and practices shall include, but are not limited to, the use of:
(a) Analyses of reference materials to ensure the reproducibility and accuracy of test results;

(b) Systems to ensure proper functioning of instruments, equipment, reagents, and performance of laboratory personnel;

(c) Valid reagents, equipment, calibration, and systematic procedural checks;

(d) A system that ensures timely and pertinent error detection and correction for the review and reporting of test results; and

(e) Quality control materials from the Centers for Disease Control and Prevention's Newborn Screening Quality Assurance Program.

B. Adequacy of Methods and Equipment. To ensure the adequacy of test methods, test systems, and equipment, a permittee shall:

(1) Employ methods, test systems, and equipment that provide accurate and reliable test results and reports;

(2) Have sufficient and suitable equipment and instruments to perform the type and volume of tests offered and performed by the laboratory;

(3) Perform tests in a manner that provides accurate and reliable test results;

(4) Monitor test methods, instruments, and testing systems, where applicable, for the following performance characteristics:
(a) Method, instrument, and test system detection limits and cutoff levels;

(b) Precision;

(c) Accuracy;

(d) Specificity;

(e) Sensitivity;

(f) Interferences; and

(g) Other related test variables that affect test results; and

(5) Have adequate quality control systems and procedures in place to:
(a) Obtain and verify a statistically valid number of counts or measurement levels to give accurate and reliable test results for an analytical instrument, each day an instrument is used for testing; and

(b) Report results:
(i) As set forth in COMAR 10.10.06.04B;

(ii) As set forth in 42 CFR § 493.1291; and

(iii) Within the time frames required by this chapter.

C. Supplies, Reagents, and Materials.

(1) A permittee shall ensure that supplies, reagents, and other materials used in the collected, handling, testing, and storing of newborn blood-spot specimens are stored, labeled, checked for proper reactivity, and used according to the provisions of this chapter.

(2) A permittee shall:
(a) Label laboratory supplies and materials used in the testing process, including reagents, solutions, controls, and calibrators to indicate the:
(i) Identity of reagents, solutions, controls, and calibrators;

(ii) Strength or concentration;

(iii) Storage requirements;

(iv) Preparation and expiration date;

(v) Identity of the preparer; and

(vi) Other pertinent information that affects test performance;

(b) Store laboratory materials in the manner recommended by the manufacturer;

(c) Test representative samples of each lot of materials used for or in the process of testing blood-spot specimens on a regularly scheduled basis to determine that each lot of test material is capable of performing as required; and

(d) Use supplies, reagents, kits, test systems, and other materials in a manner consistent with the recommendations and instructions provided by the manufacturer.

(3) A permittee may not use supplies, reagents, kits, test systems, or other materials in the testing process that have:
(a) Exceeded the materials' established expiration dates;

(b) Substandard reactivity; or

(c) Deteriorated.

Disclaimer: These regulations may not be the most recent version. Maryland may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.