A. Beam limitation.
The useful beam shall be limited to the area of clinical interest.
(1) General purpose stationary and mobile
x-Ray systems.
(a) There shall be provided a
means for stepless adjustment of the size of the x-ray field.
(b) For those units so designed, a method
shall be provided for visually defining the perimeter of the x-ray field. The
total misalignment of the edges of the visually defined field with the
respective edges of the x-ray field along either the length or width of the
visually defined field shall not exceed 2 percent of the distance from the
source to the center of the visually defined field when the surface upon which
it appears is perpendicular to the axis of the x-ray beam.
(c) The Agency may grant an exemption on
non-certified x-ray systems to F.7.A.(1)(a) and (b) provided the registrant
makes a written application for such exemption and in that application:
(i) Demonstrates it is impractical to comply
with F.7.A.(1)(a) and (b); and
(ii)
The purpose of F.7.A.(1)(a) and (b) will be met by other methods.
(2) Additional
requirements for stationary general purpose x-ray systems. In addition to the
requirements of F.7.A.(1), all stationary general purpose x-ray systems shall
meet the following requirements:
(a) A method
shall be provided to indicate when the axis of the x-ray beam is perpendicular
to the plane of the image receptor, to align the center of the x-ray field with
respect to the center of the image receptor to within 2 percent of the SID, and
to indicate the SID to within 2 percent;
(b) The beam-limiting device shall indicate
numerically the field size in the plane of the image receptor to which it is
adjusted; and
(c) Indication of
field size dimensions and SID's shall be specified in inches and/or
centimeters, and shall be such that aperture adjustments result in x-ray field
dimensions in the plane of the image receptor which correspond to those
indicated by the beam-limiting device to within 2 percent of the SID when the
beam axis is indicated to be perpendicular to the plane of the image
receptor.
(3) X-Ray
systems designed for one image receptor size. Radiographic equipment designed
for only one image receptor size at a fixed SID shall be provided with means to
limit the field at the plane of the image receptor to dimensions no greater
than those of the image receptor, and to align the center of the x-ray field
with the center of the image receptor to within 2 percent of the SID, or shall
be provided with means to both size and align the x-ray field such that the
x-ray field at the plane of the image receptor does not extend beyond any edge
of the image receptor.
(4)
Mammography equipment standards. Only x-ray systems meeting the following
standards shall be used:
(a) System design:
The x-ray system shall be specifically designed for mammography.
(b) Image receptor: The image receptor
systems and their individual components shall be specifically designed for or
appropriate for mammography.
(c)
Radiographic systems designed for mammography shall be provided with means to
limit the useful beam such that the x-ray field at the plane of the image
receptor does not extend beyond any edge of the image receptor at any
designated SID except the edge of the image receptor designed to be adjacent to
the chest wall where the x-ray field may not extend beyond this edge by more
than 2 percent of the SID. This requirement can be met with a system, which
performs as prescribed in F.7.A.(5).
(5) Special purpose x-ray systems.
(a) Means shall be provided to limit the
x-ray field in the plane of the image receptor so that such field does not
exceed each dimension of the image receptor by more than 2 percent of the SID
when the axis of the x-ray beam is perpendicular to the plane of the image
receptor.
(b) Means shall be
provided to align the center of the x-ray field with the center of the image
receptor to within 2 percent of the SID, or means shall be provided to both
size and align the x-ray field such that the x-ray field at the plane of the
image receptor does not extend beyond any edge of the image receptor.
(c) F.7.A(5)(a) and (b) may be met with a
system that meets the requirements for a general purpose x-ray system as
specified in F.7.A(1) or, when alignment means are also provided, may be met
with either:
(i) An assortment of removable,
fixed-aperture, beam-limiting devices sufficient to meet the requirement for
each combination of image receptor size and SID for which the unit is designed
with each such device having clear and permanent markings to indicate the image
receptor size and SID for which it is designed; or
(ii) A beam-limiting device having multiple
fixed apertures sufficient to meet the requirement for each combination of
image receptor size and SID for which the unit is designed. Permanent, clearly
legible markings shall indicate the image receptor size and SID for which each
aperture is designed and shall indicate which aperture is in position for
use.
B. Radiation exposure control devices.
(1) Timers. Means shall be provided to
terminate the exposure at a preset time interval,
preset product of current and time, a preset number of pulses, or a preset
radiation exposure to the image receptor. In addition, it shall not be possible
to make an exposure when the time is set to a zero or off position if either
position is provided.
(2) Manual
Exposure Control.
(a) An x-ray control shall
be incorporated into each x-ray system such that an
exposure can be terminated by the operator at any time
except for:
(i) Exposure of one-half second or
less, or
(ii) During serial
radiography when means shall be provided to permit completion of any single
exposure of the series in process.
(b) Each x-ray control shall be located in
such a way as to meet the following requirements:
(i) Stationary x-ray systems shall be
required to have the x-ray control permanently mounted in a protected area so
that the operator is required to remain in that protected area during the
entire exposure; and
(ii) Mobile and portable x-ray systems which
are:
(a) Used for greater than one week in the
same location, i.e., a room or suite, shall meet the requirements of
F.7.B.(2)(b)(i)
(b) Used for
greater than one hour and less than one week at the same location, i.e., one
room or suite, shall meet the requirement of F.7.B.(2)(b)(ii)(a) or be provided
with a 6.5 feet (1.98m) high protective barrier which is placed at least 6
(1.83m) feet from the tube housing assembly and at least 6 (1.83m) feet from
the patient; or
(c) Used to make an
exposure(s) of a patient at the use location shall meet the requirement of
F.7.B.(2)(b)(ii)(a) or(b) or be provided with a method or x-ray control which
will permit the operator to be at least six feet (3.66m) from the tube housing
assembly during an exposure.
(iii) The x-ray control shall provide visual
indication observable at or from the operator's protected position whenever
x-rays are produced. In addition, a signal audible to the operator, if
available with the unit, shall indicate that the exposure has
terminated.
(3)
Automatic
exposure controls. When an automatic
exposure control is provided:
(a) Indication shall be made on the control
panel when this mode of operation is selected;
(b) If the x-ray tube potential is equal to
or greater than 50 kVp, the minimum exposure time for field emission equipment
rated for pulsed operation shall be equal to or less than a time interval
equivalent to two pulses;
(c) The
minimum exposure time for all equipment other than that specified in
F.7.B.(3)(b) shall be equal to or less than one-sixtieth second or a time
interval required to deliver 5 mAs, whichever is greater;
(d) Either the product of peak x-ray tube
potential, current, and exposure time shall be limited to not more than 60 kWs
per exposure, or the product of x-ray tube current and exposure time shall be
limited to not more than 600 mAs per exposure except that when the x-ray tube
potential is less than 50 kVp in which case the product of x-ray tube current
and exposure time shall be limited to not more than 2000 mAs per exposure;
and
(e) A visible signal shall
indicate when an exposure has been terminated at the limits required by
F.7.B(3)(d), and manual resetting shall be required before further
automatically timed exposures can be made.
(4) Reproducibility. With a timer setting of
0.5 seconds or less, the average exposure period (T) shall be greater than or
equal to five times the maximum exposure period (Tmax) minus the minimum
exposure (Tmin) when four timer tests are performed; i.e., T > 5 (Tmax
-Tmin).
C. Source-to-skin
distance. All mobile or portable radiographic systems shall be provided with
means to limit the source-to-skin distance to not less than 30
centimeters.
D. Exposure
reproducibility. The coefficient of variation of exposure shall not exceed 0.10
when all technique factors are held constant. This requirement shall be deemed
to have been met if, when four exposures are made at identical technique
factors, that the value of the average exposure (E) is greater than or equal to
five times the maximum exposure (Emax) minus the minimum exposure (Emin), i.e.,
E > 5 (Emax - Emin).
E.
Radiation from capacitor energy storage equipment in standby status. Radiation
emitted from the x-ray tube when the exposure switch or timer is not activated
shall not exceed a rate of 2 milliroentgens (0.516 mC/kg) per hour at 5
centimeters from any accessible surface of the diagnostic source assembly with
the beam-limiting device fully open.
F. Additional requirements applicable to
certified systems only. Diagnostic x-ray systems incorporating one or more
certified component(s) shall be required to comply with the following
additional requirement(s) which relate to that certified component(s).
(1) Reproducibility. When the equipment is
operated on an adequate power supply as specified by the manufacturer in
accordance with the requirements of applicable Federal standards, the estimated
coefficient of variation of radiation exposures shall
be no greater than 0.05, for any specific combination of selected technique
factors.
(2) Linearity. When the
equipment allows a choice of x-ray tube current settings and is operated on a
power supply as specified by the manufacturer in accordance with the
requirements of applicable federal standards, any fixed x-ray tube potential
within the range of 40 percent to 100 percent of the maximum rating, the
average ratios of
exposure to the indicated
milliampere-seconds product, i.e., mR/mAs, obtained at any two consecutive tube
current settings shall not differ by more than 0.10 times their sum,
X1-X2 < 0.10 (X1 + X2),
where X1 and X2 are the average mR/mAs values obtained at
each of two consecutive tube current settings.
(3) Accuracy. Deviation of technique factors
from indicated values shall not exceed the limits specified for that system by
its manufacturer. In the absence of the manufacturer's specifications the
deviation shall not exceed 10% of the indicated value.
(4) Beam limitation for stationary and mobile
general purpose x-ray systems.
(a) There shall
be provided a means of stepless adjustment of the size of the x-ray
field.
(b) When a light localizer
is used to define the x-ray field, it shall provide an average illumination of
not less than 160 lux or 15 footcandles at 100 centimeters or at the maximum
SID, whichever is less. The average illumination shall be based upon
measurements made in the approximate center of each quadrant of the light
field. Radiation therapy simulation systems manufactured on or after May 27,
1980, are exempt from this requirement.
(c) The edge of the light field at 100
centimeters or at the maximum SID, whichever is less, shall have a contrast
ratio, corrected for ambient lighting, of not less than 4 in the case of
beam-limiting devices designed for use on stationary equipment, and a contrast
ratio of not less than 3 in the case of beam-limiting devices designed for use
on mobile equipment. The contrast ratio is defined as I1/I2 when I1 is the
illumination 3 millimeters from the edge of the light field toward the center
of the field; and I2 is the illumination 3 millimeters from the edge of the
light field away from the center of the field. Compliance shall be determined
with a measuring instrument aperture of 1 millimeter in diameter.
(5) Beam limitation for portable
x-ray systems. Beam limitation for portable x-ray systems shall meet the beam
limitation requirements of F.7.A.(1). and F.7.F.(4).
(6) Field limitation and alignment on
stationary general purpose x-ray systems. For stationary, general purpose x-ray
systems which contain a tube housing assembly, an x-ray control, and, for those
systems so equipped, a table, all certified in accordance with
21 CFR
1020.30(c):
(a) Positive beam limitation (PBL) when
present shall function whenever all the following conditions are met:
(i) The image receptor is inserted into a
permanently mounted cassette holder;
(ii) The image receptor length and width are
each less than 50 centimeters;
(iii) The x-ray beam axis is within plus or
minus 3 degrees of vertical and the SID is 90 centimeters to 130 centimeters
inclusive; or the x-ray beam axis is within plus or minus 3 degrees of
horizontal and the SID is 90 centimeters to 205 centimeters
inclusive;
(vi) The x-ray beam axis
is perpendicular to the plane of the image receptor to within plus or minus 3
degrees;
(v) Neither tomographic
nor stereoscopic radiography is being performed; and
(vi) The PBL system has not been
intentionally overridden. This override provision is subject to
F.7.F(6)(c).
(b) Positive
beam limitation (PBL) shall prevent the production of x-rays when:
(i) Either the length or width of the x-ray
field in the plane of the image receptor differs, except as permitted by
F.7.F(6)(e) from the corresponding image receptor dimensions by more than 3
percent of the SID; or
(ii) The sum
of the length and width differences as stated in F.7.F(6)(b)(i) without regard
to sign exceeds 4 percent of the SID.
(c) If a means of overriding the positive
beam limitation system exists, that means:
(i)
Shall be designed for use only in the event of PBL system failure or if the
system is being serviced; and
(ii)
If in a position that the operator would consider it part of the operational
controls or if it is referenced in the operator's manual or in other materials
intended for the operator,
(a) Shall require
that a key be utilized to defeat the PBL;
(b) Shall require that the key remain in
place during the entire time the PBL system is overridden; and
(c) Shall require that the key or key switch
be clearly and durably labeled as follows:
FOR X-RAY FIELD LIMITATION SYSTEM
FAILURE
(d)
Compliance with F.7.F.6(b) shall be determined when the equipment indicates
that the beam axis is perpendicular to the plane of the image receptor and the
provisions of F.7.F(6)(a) are met. Compliance shall be determined no sooner
than five seconds after insertion of the image receptor.
(e) The positive beam limitation system shall
be capable of operation, at the discretion of the operator, such that the size
of the field may be made smaller than the size of the image receptor through
stepless adjustment of the field size.
(f) The positive beam limitation system shall
be designed such that if a change in image receptor does not cause an automatic
return to positive beam limitation function as described in F.7.F.(6)(b), then
any change of image receptor size or SID must cause the automatic
return.
(7) Timers. Except for dental panoramic
systems, termination of exposure shall cause automatic resetting of the timer
to its initial setting or to zero.
(8) Transmission limit for image receptor
supporting devices used for mammography. For x-ray systems manufactured after
September 5, 1978 which are designed only for mammography, the transmission of
the primary beam through any image receptor support provided with the system
shall be limited such that the exposure 5 centimeters from any accessible
surface beyond the plane of the image receptor supporting device does not
exceed 0.1 milliroentgen (25.8 nC/kg) for each activation of the tube. Exposure
shall be measured with the system operated at the minimum SID for which it is
designed. Compliance shall be determined at the maximum rate peak tube
potential for the system and at the maximum rated product of the tube current
and exposure time (mAs) for that peak tube potential. Compliance shall be
determined by measurements averaged over an area of 100 square centimeters with
no linear dimension greater than 20 centimeters.