Code of Maine Rules
10 - DEPARTMENT OF HEALTH AND HUMAN SERVICES
144 - DEPARTMENT OF HEALTH AND HUMAN SERVICES - GENERAL
Chapter 220 - Rules Relating to Radiation Protection
Part F - X-RAYS IN THE HEALING ARTS
Section 144-220-F-5 - General requirements for all diagnostic x-ray systems
Current through 2024-38, September 18, 2024
In addition to other requirements of this part, all diagnostic x-ray systems shall meet the following requirements:
A. Warning label. The control panel containing the main power switch shall bear the warning statement, legible and accessible to view: "WARNING: This x-ray unit may be dangerous to patient and operator unless safe exposure factors and operating instructions are observed."
B. Battery charge indicator. On battery-powered x-ray generators, visual means shall be provided on the control panel to indicate whether the battery is in a state of charge adequate for proper operation.
C. Leakage radiation from the diagnostic source assembly. The leakage radiation from the diagnostic source assembly measured at a distance of 1 meter in any direction from the source shall not exceed 100 milliroentgens (25.8 mC/kg) in one hour when the x-ray tube is operated at its leakage technique factors. Compliance shall be determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.
D. Radiation from components other than the diagnostic source assembly. The radiation emitted by a component other than the diagnostic source assembly shall not exceed 2 milliroentgens (0.516 mC/kg) in one hour at 5 centimeters from any accessible surface of the component when it operated in an assembled x-ray system under any conditions for which it was designed. Compliance shall be determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.
E. Beam Quality.
TABLE 1
Minimum HVL (mm of AL) |
|||
Design operating range (Kilovolts peak) |
Measured potential (kVp) |
Specified dental systems* |
All other diagnostic X-ray systems |
Below 50 --------- |
30 |
1.5 |
0.3 |
40 |
1.5 |
0.4 |
|
49 |
1.5 |
0.5 |
|
50 to 70 --------- |
50 |
1.5 |
1.2 |
60 |
1.5 |
1.3 |
|
70 |
1.5 |
1.5 |
|
Above 70 --------- |
71 |
2.1 |
2.1 |
80 |
2.3 |
2.3 |
|
90 |
2.5 |
2.5 |
|
100 |
2.7 |
2.7 |
|
110 |
3.0 |
3.0 |
|
120 |
3.2 |
3.2 |
|
130 |
3.5 |
3.5 |
|
140 |
3.8 |
3.8 |
|
150 |
4.1 |
4.1 |
|
* for any dental x-ray system designed for use with intraoral image receptors and manufactured after December 1, 1980; |
TABLE 2
Filtration Required vs. Operating Voltage |
|
Total Filtration (inherent plus added) |
|
Operating Voltage (kVp) |
(millimeters aluminum equivalent) |
Below 50 ----------------------------------- |
0.5 |
50 - 70 -------------------------------------- |
1.5 |
Above 70 ----------------------------------- |
2.5 |
F. Multiple tubes. Where two or more radiographic tubes are controlled by one exposure switch, the tube or tubes, which have been selected, shall be clearly indicated prior to initiation of the exposure. This indication shall be on the x-ray control panel.
G. Mechanical support of tube head. The tube housing assembly supports shall be adjusted such that the tube housing assembly will remain stable during an exposure unless tube housing movement is a designed function of the x-ray system.
H. Technique indicators.
I. Quality assurance program