Section 144-220-F-12 - Bone densitometry

Current through 2024-38, September 18, 2024

In addition to other requirements of this part, all x-ray bone densitometry systems shall meet the following requirements:

A. The requirement for having a shielding design performed by a qualified expert for bone densitometry units has been waived.

B. An inspection of the unit and an environmental survey of the area shall be performed, within 30 days after installation to ensure that radiation levels to members of the general public outside the exam room are within regulatory limits. A copy of this report must be sent to the Maine Center for Disease Control and Prevention, Radiation Control Program.

C. The console shall be positioned such that the operator is at least 1 meter from the scan field.

D. Access to the exam room during a scan shall be controlled by the operator to prevent exposure to other personnel and the general public.

Disclaimer: These regulations may not be the most recent version. Maine may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.