Code of Maine Rules
10 - DEPARTMENT OF HEALTH AND HUMAN SERVICES
144 - DEPARTMENT OF HEALTH AND HUMAN SERVICES - GENERAL
Chapter 220 - Rules Relating to Radiation Protection
Part F - X-RAYS IN THE HEALING ARTS
Section 144-220-F-12 - Bone densitometry
Current through 2024-38, September 18, 2024
In addition to other requirements of this part, all x-ray bone densitometry systems shall meet the following requirements:
A. The requirement for having a shielding design performed by a qualified expert for bone densitometry units has been waived.
B. An inspection of the unit and an environmental survey of the area shall be performed, within 30 days after installation to ensure that radiation levels to members of the general public outside the exam room are within regulatory limits. A copy of this report must be sent to the Maine Center for Disease Control and Prevention, Radiation Control Program.
C. The console shall be positioned such that the operator is at least 1 meter from the scan field.
D. Access to the exam room during a scan shall be controlled by the operator to prevent exposure to other personnel and the general public.