A. The ASC shall have policies and procedures, approved and implemented by the medical staff and governing body, that address provisions for:
B. Policies and procedures shall be developed, implemented and approved by the ASCs governing body for the types and numbers of sterilizing equipment and autoclaves sufficient to meet the surgical sterilization needs of the ASC.
C. All steam sterilizing equipment shall have live bacteriological spore monitoring performed at a frequency according to the manufacturers instructions.
D. All ethylene oxide sterilizing equipment shall have live bacteriological spore monitoring performed with each load and according to manufacturers recommendation. There shall be ventilation of the room used for this sterilization to the outside atmosphere. There shall be a system in place to monitor trace gases of ethylene oxide with a working alert system which is tested and documented daily.