A. The DME provider shall not furnish any legend device or medical gas to a patient without a prescription or medical order from a licensed practitioner with prescriptive authority.
B. General Requirements
1. The provider shall establish
a suitable facility to house the equipment, allow for equipment maintenance
work space, and contain sufficient space for the storage and retrieval of all
required records.
2. The provider
shall maintain the facility in a clean, orderly and sanitary condition at all
times.
3. The facility shall be
equipped with a functioning lavatory with hot and cold running water, or in the
alternative, hand washing appliances or waterless hand cleaner are
available.
4. The facility shall
comply with all local and state building laws and fire codes.
5. The provider shall comply with all
requirements from the United States Pharmacopeia (USP), the federal Food and
Drug Administration (FDA), federal Department of Transportation (DOT) and
Occupational Safety and Health Administration (OSHA) relative to the storage,
packaging, labeling and shipping of DME including medical gases.
6. The provider shall staff the facility with
an adequate number of qualified personnel to properly render DME services in
the manner prescribed by law.
7.
The provider shall make services continuously available without interruption
when such services are essential to the maintenance of life or when the lack of
services might reasonably cause harm.
8. The provider shall implement and maintain
written procedures for handling complaints, and further, shall maintain a
complaint file documenting all complaints and their resolution.
C. Requirements for Providers of Medical Gas, Oxygen and Respiratory Equipment
1. The provider shall comply with the
following:
a. when transporting medical gas or
oxygen in cylinder or liquid form, comply with all current DOT rules;
b. when trans-filling medical oxygen systems,
comply with FDA and all state agency requirements regarding trans-filling and
repackaging;
c. demonstrate that
medical gas and oxygen provided in cylinder or liquid form meet minimum purity
standards for medical grade gas or medical grade oxygen; and
d. adhere to the following safety inspection
requirements:
i. demonstrate that each piece
of oxygen or respiratory equipment has been checked, is free of defects, and
operates within the manufacturer's specifications;
ii. refrain from modifying equipment to the
extent that the modification might reasonably cause harm;
iii. maintain all electrical components so
they do not present fire or shock hazard; and
iv. ensure that all appropriate warning
labels or labeling, including tags, are present on the equipment
provided.
2.
The provider shall comply with the following recall procedures:
a. ensure that lot numbers and expiration
dates are affixed to each cylinder delivered;
b. maintain a tracking system for all medical
gas and oxygen delivered;
c.
document all equipment serial numbers and model numbers to ensure that
equipment can be retrieved in the event a recall is initiated; and
d. maintain records for equipment that
requires FDA tracking.
3. The provider shall comply with the
following maintenance and cleaning requirements:
a. maintain documentation demonstrating that
a function and safety check of equipment was performed prior to
set-up;
b. maintain an established
protocol for cleaning and disinfecting equipment which addresses both aerobic
and anaerobic pathogens;
c.
maintain a material safety data sheet (MSDS) on file for solutions and products
used in cleaning and disinfecting procedures;
d. maintain segregated areas on the premises
and in delivery vehicles for clean, dirty and contaminated equipment;
e. clean and disinfect equipment according to
manufacturers' specifications;
f.
instruct the patient or caregiver on proper cleaning techniques as specified by
the manufacturer; and
g. ensure
that all medical gas, oxygen and respiratory equipment is properly identified
by a tag or label as to its current status of use, i.e., out-of-order or ready
for use.
4. The provider
shall implement a comprehensive preventive maintenance program which shall
include the following:
a. procedures for
problem reporting, tracking, recall, and resolution;
b. performance of service as specified by the
manufacturer and the documentation of such performance in the service records;
and
c. routine inspection, service,
and maintenance of equipment located in the patient's home according to the
manufacturer's specifications.
5. The provider shall maintain repair logs to
document repair and maintenance of equipment, and such logs shall contain the
following information:
a. type of
equipment;
b.
manufacturer;
c. model;
d. serial number;
e. date of repair;
f. specific repair made; and
g. name of person or company performing the
repair.
6. The provider
shall maintain testing equipment to ensure accurate calibration. Testing
equipment shall be appropriate for the level of service offered. Scales used to
weigh liquid oxygen reservoirs shall be properly maintained to ensure
accuracy.
7. The provider shall
utilize client orientation checklists to review the following information with
the patient or care giver:
a. instructions
for use of the equipment;
b. safety
precautions;
c. cleaning
procedures;
d. maintenance
procedures;
e. return
demonstrations on back-up oxygen systems delivered;
f. instruction for emergency and routine
contact procedures; and
g. delivery
and review of written instruction materials to ensure the patient receives
adequate information to properly operate the equipment.
8. A written plan of service shall be
developed, implemented, and documented in the patient record. The plan of
service shall include, but is not limited to, an assessment of the safety of
the home environment, the ability of the patient or care giver to comply with
the prescription or medical order, and the ability of the patient or care giver
to operate and clean the equipment as instructed.
D. Requirements for Providers of Other Durable Medical Equipment
1. Providers who
sell, rent or furnish other DME or legend devices shall comply with the
following:
a. provide proper training to
personnel for the safe delivery and use of any DME or legend device;
and
b. ensure that all
manufacturer's recommended assembly and maintenance procedures are followed;
and
c. adhere to the following
safety inspection measures:
i. demonstrate
that each piece of DME or legend device has been checked, is free of defect and
operates within the manufacturer's specifications;
ii. refrain from modifying equipment to the
extent that the modification might reasonably cause harm;
iii. maintain all electrical components so
they do not present fire or shock hazard; and
iv. ensure that all appropriate warning
labels or labeling, including tags, are present on the equipment
provided.
2.
The provider shall comply with the following maintenance and cleaning
requirements:
a. maintain documentation
demonstrating that a function and safety check of equipment was performed prior
to set-up;
b. maintain an
established protocol for cleaning and disinfecting equipment which addresses
both aerobic and anaerobic pathogens;
c. maintain a material safety data sheet
(MSDS) on file for solutions and products used in cleaning and disinfecting
procedures;
d. maintain segregated
areas on the premises and in delivery vehicles for clean, dirty and
contaminated equipment;
e. clean
and disinfect equipment according to manufacturers' specifications;
and
f. instruct the patient or
caregiver on proper cleaning techniques as specified by the
manufacturer.
E. Records Management for All DME Providers
1. An electronic record keeping system shall
be implemented and maintained by the provider. The system shall provide
adequate safeguards against unauthorized access, manipulation or alternation,
and further, shall be susceptible to reconstruction in the event of electronic
or computer malfunction or an unforeseen accident resulting in the destruction
of the system or the information contained therein.
2. All records required in this Chapter shall
be retained for a minimum of two years from the last transaction.
3. All records required in this Chapter shall
be available and readily retrievable upon request for board inspection and
review. In particular, such records shall be produced within 72 hours of the
request.
AUTHORITY NOTE: Promulgated in accordance with R.S. 37:1182.
Disclaimer: These regulations may not be the most recent version. Louisiana may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.