Section 902 KAR 100:130 - Dental
Current through Register Vol. 51, No. 3, September 1, 2024
RELATES TO: KRS 211.842-211.852, 211.990(4)
NECESSITY, FUNCTION, AND CONFORMITY: The Cabinet for Human Resources is authorized by KRS 211.844 to provide by administrative regulation for the registration and licensing of the possession or use of any sources of ionizing or electronic product radiation and the handling and disposal of radioactive waste. The purpose of this administrative regulation is to provide special requirements for the possession, use and operation of intra-oral dental radiographic x-ray systems.
Section 1. Applicability. This administrative regulation shall apply to dental intra-oral radiographic x-ray systems and to persons, equipment and materials used in connection with the possession, use or operation of these systems.
Section 2. Source to Skin Distance. Each radiographic x-ray system designed for use with an intra-oral image receptor shall be provided with a means to limit the source to skin distance to not less than:
Section 3. Field Limitation. Each radiographic x-ray system designed for use with an intra-oral image receptor shall be provided with a means to limit the x-ray beam.
Section 4. Operator Protection. Each installation shall be provided with a protective barrier for the operator or shall be so arranged that the operator can conveniently stand, in the judgment of the cabinet, at least one and eight tenths (1.8) meters from the patient, the tube housing assembly, and the useful beam if the exposure to the operator is within the limits provided by 902 KAR 100:020, Section 20.
Section 5. Operating Procedures. In performing intra-oral dental radiography the following rules shall apply:
Section 6. Filtration. In addition to the requirements of 902 KAR 100:115, Section 6, all intra-oral dental radiographic systems manufactured on and after December 1, 1980, shall have a minimum half-value layer not less than one and five- tenths (1.5) millimeters aluminum equivalent filtration permanently installed in the useful beam.
Section 7. Linearity. On dental intra-oral radiographic systems certified under the federal performance standard, if the equipment allows a choice of x-ray tube current settings and is operated on a power supply as specified by the manufacturer in accordance with the requirements of applicable federal standards, for any fixed x-ray tube potential within the range of forty (40) to 100 percent of the maximum rating, the average ratios of exposure to the indicated milliampere-seconds product obtained at any two (2) consecutive tube current settings shall not differ by more than one-tenth (0.1) times their sum.
STATUTORY AUTHORITY: KRS 194.050, 211.090, 211.844