Section 67-7-4 - Calibration of audiometric equipment
Universal Citation: KS Admin Regs 67-7-4
Current through Register Vol. 43, No. 39, September 26, 2024
(a) Any audiometric equipment that is used in connection with the sale or fitting of hearing aids in this state shall be calibrated at intervals not exceeding two years.
(b) Each licensee shall submit to the board written proof of each calibration required in subsection (a) of this rule and regulation within 30 days of the required dates of calibration. Written proof of calibration shall include the following:
(1) The name of the owner;
(2) the make, model, and serial number of the
equipment;
(3) the date of
calibration;
(4) the printed name
of the person and company calibrating the equipment;
(5) the signature of the person calibrating
the equipment; and
(6) the name of
the licensee submitting proof of calibration.
(c) Any of the following acts may constitute unethical conduct for which a licensee may be disciplined under K.S.A. 74-5818 and amendments thereto:
(1) Provides false or
misleading information under this rule and regulation;
(2) uses audiometric equipment that has not
been calibrated as required by this rule and regulation; or
(3) uses audiometric equipment the
calibration of which has not been reported as required by this rule and
regulation.
Disclaimer: These regulations may not be the most recent version. Kansas may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.