Section 28-35-345 - Radiation survey instruments
Current through Register Vol. 43, No. 52, December 26, 2024
(a) Each licensee or registrant shall maintain a sufficient supply of calibrated and operable radiation survey instruments capable of detecting beta and gamma radiation at each field station and temporary job site to make any physical radiation survey required by this part and part 4 of these regulations. Instrumentation shall be capable of measuring a range of 0.1 milliroentgens through at least 50 milliroentgens per hour.
(b) Each licensee shall have available additional calibrated and operable radiation-detection instruments sensitive enough to detect the low radiation and contamination levels that could be encountered if a sealed source ruptures. Any licensee may own the instruments or may have a procedure to obtain them from a second party when needed.
(c) Within the previous six months and after each servicing or repair, each radiation survey instrument used shall be calibrated as follows:
(d) A calibration record for each instrument shall be maintained for two years for inspection by the department.