Section 28-35-315 - Particle accelerator controls and interlock systems

Universal Citation: KS Admin Regs 28-35-315

Current through Register Vol. 43, No. 52, December 26, 2024

(A) Instrumentation, readouts and controls on the particle accelerator control console shall be clearly identified and easily discernible.

(B) All entrances into a target room or other high radiation area shall be provided with interlocks that shut down the machine under conditions of barrier penetration.

(C) When an interlock system has been tripped, it shall only be possible to resume operation of the accelerator by manually resetting controls at the position where the interlock has been tripped, and lastly at the main control console.

(D) Each safety interlock shall be on a circuit which shall allow its operation independently of all other safety interlocks.

(E) All safety interlocks shall be fail safe, i.e., designed so that any defect or component failure in the interlock system prevents operation of the accelerator.

(F) A scram button or other emergency power cutoff switch shall be located and easily identifiable in all high radiation areas. Such a cutoff switch shall include a manual reset so that the accelerator cannot be restarted from the accelerator control console without resetting the cutoff switch.

Disclaimer: These regulations may not be the most recent version. Kansas may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.