Section 28-35-275 - Limits on levels of radiation for radiographic exposure devices and storage containers

Universal Citation: KS Admin Regs 28-35-275
Current through Register Vol. 43, No. 52, December 26, 2024

Radiographic exposure devices measuring less than four inches from the sealed source storage position to any exterior surface of the device shall have no radiation level in excess of 50 milliroentgens per hour at six inches from any exterior surface of the device. Radiographic exposure devices measuring four or more inches from the sealed source storage position to any exterior surface of the device, and all storage containers for sealed sources or outer containers for radiographic exposure devices, shall have no radiation level in excess of 200 milliroentgens per hour at any exterior surface, or in excess of 10 milliroentgens per hour at one meter from any exterior surface. The radiation level emanating from a device or container shall be measured with the sealed source in the shielded position.

Disclaimer: These regulations may not be the most recent version. Kansas may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.