Current through Register Vol. 43, No. 52, December 26, 2024
(a) Beam
limitation, except for mammographic systems. Each registrant shall ensure that
the useful beam is collimated to the area of clinical interest. This
requirement shall be deemed to have been met if a positive beam-limiting device
meeting the manufacturer's specifications and the requirements of this
regulation has been used or if evidence of collimation is shown on at least
three sides or three corners of the film, including projections from the
shutters of the collimator, cone cutting at the corners, and borders at the
film's edge.
(1) General-purpose stationary
and mobile X-ray systems including veterinary systems, other than portable
systems, installed after the effective date of these regulations.
(A) Each registrant shall use only X-ray
systems provided with the means for independent stepless adjustment of at least
two dimensions of the X-ray field.
(B) Each registrant shall ensure that a
method is provided for visually defining the perimeter of the X-ray field. The
total misalignment of the edges of the visually defined field with the
respective edges of the X-ray field along either the length or width of the
visually defined field shall not exceed two percent of the distance from the
source to the center of the visually defined field when the surface upon which
the visually defined field appears is perpendicular to the axis of the X-ray
beam.
(C) An exemption from
paragraphs (a)(1)(A) and (B) may be granted by the secretary for noncertified
X-ray systems if the registrant submits a written application for the exemption
and that application meets the following conditions:
(i) Demonstrates that it is impractical to
comply with paragraphs (a)(1)(A) and (B); and
(ii) demonstrates that the purpose of
paragraphs (a)(1)(A) and (B) will be met by other methods.
(2) Additional
requirements for stationary general-purpose X-ray systems. In addition to the
requirements of paragraph (a)(1), each registrant shall ensure that all
stationary general-purpose X-ray systems, both certified and noncertified, meet
all of the following requirements:
(A) A
method shall be provided to indicate when the axis of the X-ray beam is
perpendicular to the plane of the image receptor, to align the center of the
X-ray field with respect to the center of the image receptor to within two
percent of the SID, and to indicate the SID to within two percent.
(B) The beam-limiting device shall indicate
numerically the field size in the plane of the image receptor to which the
device is adjusted.
(C) The
indication of the field size dimensions and SID shall be specified in inches or
centimeters, or both, and shall be such that aperture adjustments result in
X-ray field dimensions in the plane of the image receptor that correspond to
X-ray field dimensions indicated by the beam-limiting device to within two
percent of the SID when the beam axis is indicated to be perpendicular to the
plane of the image receptor.
(3) X-ray systems designed for one size of
image receptor. Radiographic equipment designed for only one size of image
receptor at a fixed SID shall be provided with the means to limit the field at
the plane of the image receptor to dimensions no greater than those of the
image receptor and to align the center of the X-ray field with the center of
the image receptor to within two percent of the SID, or shall be provided with
the means to both adjust the size of and align the X-ray field so that the
X-ray field at the plane of the image receptor does not extend beyond any edge
of the image receptor.
(4) X-ray
systems other than those described in this subsection and veterinary systems
installed before the effective date of this regulation and all portable
veterinary X-ray systems.
(A) A means shall
be provided to limit the X-ray field in the plane of the image receptor so that
the field does not exceed each dimension of the image receptor by more than two
percent of the SID when the axis of the X-ray beam is perpendicular to the
plane of the image receptor.
(B) A
means shall be provided to align the center of the X-ray field with the center
of the image receptor to within two percent of the SID, or the means shall be
provided to both adjust the size of and align the X-ray field so that the X-ray
field at the plane of the image receptor does not extend beyond any edge of the
image receptor. Compliance shall be determined with the axis of the X-ray beam
perpendicular to the plane of the image receptor.
(C) The requirements of paragraphs (a)(4)(A)
and (B) may be met with a system that meets the requirements for a
general-purpose X-ray system as specified in paragraph (a)(1) or, when
alignment means are also provided, may be met with either of the following:
(i) An assortment of removable,
fixed-aperture, beam-limiting devices sufficient to meet the requirements for
each combination of image receptor size and SID for which the system is
designed, with each such device having clear and permanent markings to indicate
the image receptor size and SID for which the system is designed; or
(ii) a beam-limiting device having multiple
fixed apertures sufficient to meet the requirements for each combination of
image receptor size and SID for which the system is designed. Permanent,
clearly legible markings shall indicate the image receptor size and SID for
which each aperture is designed and shall indicate which aperture is in
position for use.
(b) Radiation exposure control.
(1) Exposure initiation. A means shall be
provided to initiate the radiation exposure by a deliberate action on the part
of the operator, including the depression of a switch. Radiation exposure shall
not be initiated without such an action. In addition, it shall not be possible
to initiate an exposure when the timer is set to a "zero" or "off" position if
either position is provided.
(2)
Exposure indication. A means shall be provided for visual indication observable
at or from the operator's protected position whenever X-rays are produced. In
addition, a signal audible to the operator shall indicate that the exposure has
terminated.
(3) Exposure
termination. A means shall be provided to terminate the exposure at a preset
time interval, preset product of current and time, a present number of pulses,
or a preset radiation exposure to the image receptor. Except for dental
panoramic systems, termination of each exposure shall cause the automatic
resetting of the timer to its initial setting or to "zero."
(A) Manual exposure control. An X-ray control
shall be incorporated into each X-ray system so that each exposure can be
terminated by the operator at any time, except for either of the following:
(i) During any exposure of one-half second or
less; or
(ii) during serial
radiography, when a means shall be provided to permit the completion of any
single exposure of the series in process.
(B) Automatic exposure controls. When an
automatic exposure control is provided, all of the following requirements shall
be met:
(i) The mode of operation selected
shall be indicated on the control panel.
(ii) If the X-ray tube potential is equal to
or greater than 50kVp, the minimum exposure time for field emission equipment
rated for pulsed operation shall be equal to or less than a time interval
equivalent to two pulses.
(iii)
The minimum exposure time for all equipment other than that specified in
paragraph (b)(2)(B) shall be equal to or less than one-sixtieth of a second or
the time interval required to deliver 5 mAs, whichever is greater.
(iv) Either the product of the peak X-ray
tube potential, current, and exposure time shall be limited to not more than 60
kWs per exposure, or the product of the X-ray tube current and exposure time
shall be limited to not more than 600 mAs per exposure. However, when the X-ray
tube potential is less than 50 kVp, the product of the X-ray tube current and
exposure time shall be limited to not more than 2,000 mAs per exposure.
(v) A visible signal shall
indicate when an exposure has been terminated at the limits required by
paragraph (b)(2)(D), and manual resetting shall be required before further
automatically timed exposures can be made.
(4) Exposure duration or timer linearity. For
systems that provide for the independent selection of exposure time settings,
the average ratios (X1) of exposure to the indicated timer setting, in units of
C kg-1s-1 (mR/s), obtained at any two clinically used timer settings shall not
differ by more than 0.10 times their sum. This calculation is written as
follows:
(X-1 -X2) < 0.1 (X-1 + X2)
where X-1 and X2 are the average C kg-1 s-1 (mR/s) values.
(5) Exposure control
location. The X-ray exposure control shall be placed so that the operator can
view the patient while making any exposure.
(6) Operator protection, except for
veterinary systems.
(A) Stationary systems.
Each stationary X-ray system shall be required to have the X-ray control
permanently mounted in a protected area so that the operator is required to
remain in that protected area during the entire exposure.
(B) Mobile and portable systems. Each mobile
and each portable X-ray system shall meet the following requirements:
(i) If used for one week or more at the same
location, including a room or a suite, meet the requirements of paragraph
(b)(6)(A); or
(ii) if used for
less than one week at the same location, be provided with either a protective
barrier that is at least two meters or 6.5 feet high for operator protection
during exposures or a means to allow the operator to be at least 2.7 meters or
9 feet from the tube housing assembly during the exposure.
(7) Operator protection
for veterinary systems. All stationary, mobile, or portable X-ray systems used
for veterinary work shall be provided with either a protective barrier that is
at least two meters or 6.5 feet high for operator protection during exposures
or a means to allow the operator to be at least 2.7 meters or nine feet from
the tube housing assembly during exposures.
(c) Source-to-skin distance. Each mobile or
portable radiographic system shall be provided with the means to limit the
source-to-skin distance to equal to or greater than 30 centimeters, except for
veterinary systems or systems specifically approved by the FDA.
(d) Exposure reproducibility. When all
technique factors are held constant, including control panel selections
associated with automatic exposure control systems, the coefficient of
variation of exposure for both manual and automatic exposure control systems
shall not exceed 0.05. This requirement shall apply to all clinically used
techniques.
(e) Radiation from
capacitor energy storage equipment in standby status. The radiation emitted
from the X-ray tube when the system is fully charged and the exposure switch or
timer is not activated shall not exceed a rate of 0.5 µC/kg (2
milliroentgens) per hour at five centimeters from any accessible surface of the
diagnostic source assembly, with the beam-limiting device fully open.
(f) Accuracy. Deviation of the
measured technique factors from the indicated values of kVp and exposure time
shall not exceed the limits specified for each system by its manufacturer. In
the absence of manufacturer's specifications, the deviation shall not exceed 10
percent of the indicated value for kVp and 20 percent for exposure time.
(g) mA and mAs linearity. The
following requirements shall apply when the equipment is operated with a power
supply as specified by the manufacturer for any fixed X-ray tube potential
within the range of 40 percent to 100 percent of the maximum rated mA or mAs.
(1) Equipment that provides for the
independent selection of X-ray tube current (mA). The average ratios (X-1) of
exposure to the indicated milliampere-seconds product C kg-1 mAs-1 (or mR/mAs)
obtained at any two consecutive tube current settings shall not differ by more
than 0.10 times their sum. This calculation is written as follows:
(X-1-X2) < 0.10 (X-1 + X2)
where X-1 and X2 are the average values obtained at each of two
consecutive tube current settings, or at two settings differing by no more than
a factor of two when the tube current selection is continuous.
(2) Equipment that has a
combined X-ray tube current-exposure time product (mAs) selector, but not a
separate tube current (mA) selector. The average ratios (X-1) of exposure to
the indicated milliampere-seconds product, in units of C kg-1 mAs-1 (or
mR/mAs), obtained at any two consecutive mAs selector settings shall not differ
by more than 0.10 times their sum. This calculation is written as follows:
(X-1-X2) < 0.10 (X-1 + X2)
where X-1 and X2 are the average values obtained at each of two
mAs selector settings, or at two settings differing by no more than a factor of
two when the mAs selector provides continuous selection.
(3) Measuring compliance. Determination of
compliance shall be based on 10 exposures taken within one hour, at each of the
two settings specified by this subsection. These two settings may include any
two focal spot sizes, unless one is equal to or less than 0.45 millimeters and
the other is greater than 0.45 millimeters. For purposes of this requirement,
focal spot size shall mean the nominal focal spot size specified by the X-ray
tube manufacturer.
(h)
Additional requirements applicable to certified systems only. Diagnostic X-ray
systems incorporating one or more certified components shall be required to
meet the following additional requirements that relate to that certified
component or components.
(1) Beam limitation
for stationary and mobile general-purpose X-ray systems.
(A) A means of stepless adjustment of the
size of the X-ray field shall be provided. The minimum field size at an SID of
100 centimeters shall be equal to or less than five centimeters by five
centimeters.
(B) If a light
localizer is used to define the X-ray field, the localizer shall provide an
average illumination of not less than 160 lux or 15 foot-candles at 100
centimeters or at the maximum SID, whichever is less. The average illumination
shall be based on measurements made in the approximate center of each quadrant
of the light field. Radiation therapy simulation systems manufactured on and
after May 27, 1980 shall be exempt from this requirement.
(C) The edge of the light field at 100
centimeters or at the maximum SID, whichever is less, shall have a contrast
ratio, when corrected for ambient lighting, of not less than four for
beam-limiting devices designed for use on stationary equipment and a contrast
ratio of not less than three for beam-limiting devices designed for use on
mobile equipment. The contrast ratio shall be defined as I1/I2 where I1 is the
illumination at three millimeters from the edge of the light field toward the
center of the field, and I2 is the illumination at three millimeters from the
edge of the light field away from the center of the field. Compliance shall be
determined by using a measuring instrument that has an aperture of one
millimeter in diameter.
(2) Beam limitation and alignment on
stationary general-purpose X-ray systems equipped with positive beam limitation
(PBL). If PBL is being used, all of the following requirements shall be met:
(A) The PBL shall prevent the production of
X-rays when either of the following occurs:
(i) The length or width of the X-ray field in
the plane of the image receptor differs, except as permitted by paragraph
(h)(2)(C), from the corresponding image receptor dimensions by more than three
percent of the SID.
(ii) The sum
of the length and width differences as stated in paragraph (h)(2)(A)(i),
without regard to sign, exceeds four percent of the SID.
(B) Compliance with paragraph (h)(2)(A) shall
be determined when the equipment indicates that the beam axis is perpendicular
to the plane of the image receptor. Compliance shall be determined no sooner
than five seconds after insertion of the image receptor.
(C) The PBL system shall be capable of
operation, at the discretion of the operator, so that the size of the field can
be made smaller than the size of the image receptor through stepless adjustment
of the field size. The minimum field size at an SID of 100 centimeters shall be
equal to or less than five centimeters by five centimeters.
(D) The PBL system shall be designed so that
if a change in image receptor does not cause an automatic return to PBL
function as described in paragraph (h)(2)(A), then any change of image receptor
size or SID shall cause the automatic return to PBL function.
(3) Beam limitation for portable
X-ray systems. Beam limitation for each portable X-ray system shall meet the
beam limitation requirements in paragraph (a)(1) or (h)(2).
(i) Tube stands for portable X-ray
systems. A tube stand or other mechanical support shall be used for portable
X-ray systems, so that the X-ray tube housing assembly does not need to be
handheld during exposures.