Sec. 4.

All patient care equipment must be in good working order and regularly serviced and maintained as follows:

(1) All patient care equipment must be on a documented maintenance schedule of appropriate frequency in accordance with one (1) of the following:

(A) Acceptable standards of practice.

(B) The manufacturer's recommended maintenance schedule.

(2) There must be evidence of preventive maintenance on all patient care equipment.

(3) Appropriate records must be:

(A) kept pertaining to equipment maintenance, repairs, and electrical current leakage checks; and

(B) analyzed at least triennially.

(4) Defibrillators, if present, must be discharged at least in accordance with manufacturers' recommendations, and a discharge log with initialed entries must be maintained.

Disclaimer: These regulations may not be the most recent version. Indiana may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.