Indiana Administrative Code
Title 327 - WATER POLLUTION CONTROL DIVISION
Article 8 - PUBLIC WATER SUPPLY
Rule 2 - Drinking Water Standards
Section 2-8.7 - Analytical and monitoring requirements; fecal coliform, total coliform, turbidity, disinfection for public water systems
Current through September 18, 2024
Authority: IC 13-13-5; IC 13-14-8-7; IC 13-14-9; IC 13-18-3; IC 13-18-16
Affected: IC 13-11-2; IC 13-14-8; IC 13-18-1; IC 13-18-2
Sec. 8.7.
Only the analytical methods and procedures specified in this section, or otherwise approved by EPA, may be used to demonstrate compliance by public water systems with sections 8.5 and 8.6 of this rule. Measurements for residual disinfectant concentrations must be conducted using methods specified in this rule. Analysis of pH and temperature must be conducted in accordance with one (1) of the methods listed at 40 CFR 141.23(k)(1). Analysis of total coliforms, fecal coliforms, heterotrophic bacteria, and turbidity must be conducted in accordance with one (1) of the following analytical methods or one (1) of the alternate methods listed in Appendix A to Subpart C of 40 CFR 141, and by using analytical test procedures contained in Technical Notes on Drinking Water Methods, EPA-600/R-94-173, October 1994. This document is available from the National Service Center for Environmental Publications (NSCEP), P.O. Box 42419, Cincinnati, OH 45242-0419 or http://www.epa.gov/nscep/. Measurements for total coliforms, fecal coliforms, and HPC must be conducted by a laboratory certified by the commissioner or EPA under 40 CFR 141.28 *. Until laboratory certification criteria are developed for the analysis of fecal coliforms and HPC, any laboratory certified for total coliforms analysis by the commissioner or EPA is deemed certified for fecal coliforms and HPC analysis. The following procedures shall be conducted in accordance with the publications listed as follows:
(1) Total coliform1 as set forth in the following:
(2) Fecal coliforms1 as set forth in:
(3) Heterotrophic bacteria1, Method 9215B*, pour plate method.
(4) Turbidity as set forth in:
(5) Residual disinfectant concentrations for free chlorine and combined chlorine (chloramines) as set forth in the following methods or with the alternative methods listed in Appendix A to Subpart C of 40 CFR 141. If approved by the commissioner, residual disinfectant concentrations for free chlorine and combined chlorine also may be measured using DPD colorimetric test kits. In addition, the commissioner may approve the use of the ITS free chlorine test strip for the determination of free chlorine. Use of the test strips is described in Method D99-003, "Free Chlorine Species (HOCl- and Ocl-) by Test Strip", Revision 3.0, November 21, 2003, available from Industrial Test Systems, Inc., 1875 Langston St., Rock Hill, SC 29730. Free and total chlorine residuals may be measured continuously by adapting a specific chlorine residual method for use with a continuous monitoring instrument provided the chemistry, accuracy, and precision remain the same. Instruments used for continuous monitoring must be calibrated with a grab sample measurement at least every five (5) days, or with a protocol approved by the commissioner:
(6) Residual disinfectant concentrations for ozone by the indigo method, Method 4500-O3 B*.
(7) Residual disinfectant concentrations for chlorine dioxide must be measured by Method 4500-ClO2 C, amperometric method, Method 4500-ClO2 E*, amperometric method, or Method 4500-ClO2 D*, DPD method.
(8) Residual disinfectant concentrations for total chlorine by the following methods:
*The following methods are incorporated by reference:
These methods are available for copying at the Indiana Department of Environmental Management, Office of Water Quality, 100 North Senate Avenue, Room N1255, Indianapolis, Indiana 46204.
1The time from sample collection to initiation of analysis may not exceed thirty (30) hours. Systems must hold samples below ten (10) degrees Celsius during transit.
2Lactose broth, as commercially available, may be used instead of lauryl tryptose broth if the system conducts at least twenty-five (25) parallel tests between this medium and lauryl tryptose broth using the water normally tested, and this comparison demonstrates that the false-positive rate and false-negative rate for total coliforms using lactose broth, is less than ten percent (10%).
3Media should cover inverted tubes at least one-half (1/2) to two-thirds (2/3) after the sample is added.
4No requirement exists to run the completed phase on ten percent (10%) of all total coliform-positive confirmed tubes.
5The ONPG-MUG test is also known as the Autoanalysis Colilert System.
6MI Agar may also be used.
7Six (6) times formulation strength may be used if the medium is filter-sterilized rather than autoclaved.
8The Colisure test may be read after an incubation time of twenty-four (24) hours.
9A-1 broth may be held up to three (3) months in a tightly closed screwcap tube at four (4) degrees Celsius.