Indiana Administrative Code
Title 260 - STATE DEPARTMENT OF TOXICOLOGY
Article 2.5 - BREATH TEST OPERATORS AND INSTRUMENTS
Rule 3 - Selection, Inspection, and Certification of Breath Test Instruments and Chemicals
Section 3-2 - Inspection of breath test instruments

Universal Citation: 260 IN Admin Code 3-2

Current through March 20, 2024

Authority: IC 9-30-6-5

Affected: IC 9-30-6-5

Sec. 2.

(a) A person authorized by the department shall inspect each breath test instrument deployed for evidentiary use at the instrument's established location at least once every one hundred eighty (180) days. If the location of a breath test instrument is changed, the instrument must be inspected and certified under this rule prior to use for evidentiary testing.

(b) The inspection shall include at least one (1) test demonstrating that the breath test instrument:

(1) is in good operating condition; and

(2) satisfies the accuracy requirements in subsection (e).

(c) The inspection shall include tests using reference materials certified to contain a specific concentration of ethanol with a measurement uncertainty at a stated level of confidence.

(d) The numerical analytical results of inspection tests shall be expressed to the third decimal place.

(e) The numerical analytical results of Intox EC/IR II breath test instruments shall not deviate more than five percent (5%) or five-thousandths (0.005) grams per two hundred ten (210) liters, whichever is greater, from the value of the reference material or the value of the reference material as adjusted for the ambient barometric pressure.

Disclaimer: These regulations may not be the most recent version. Indiana may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.