Illinois Administrative Code
Title 32 - ENERGY
Part 351 - RADIATION SAFETY REQUIREMENTS FOR WIRELINE SERVICE OPERATIONS AND SUBSURFACE TRACER STUDIES
Subpart C - EQUIPMENT
Section 351.330 - Radiation Detection Instruments
Current through Register Vol. 48, No. 12, March 22, 2024
a) The licensee or registrant shall keep sufficient calibrated and operable radiation survey instruments at each field station and temporary jobsite to make physical radiation surveys as required by this Part and by 32 Ill. Adm. Code 340.510(a). Instrumentation shall be capable of measuring 0.001 mSv (0.1 mrem) per hour through at least 0.5 mSv (50 mrem) per hour.
b) The licensee shall have available additional calibrated and operable radiation detection instruments sensitive enough to detect the low radiation and contamination levels that could be encountered if a sealed source ruptured. The licensee may own the instruments or may have a procedure to obtain them quickly from a second party.
c) The licensee must have each radiation survey instrument required under subsections (a) and (b) calibrated:
d) The licensee shall retain calibration records for a period of 3 years after the date of calibration for inspection by the Agency.