Illinois Administrative Code
Title 32 - ENERGY
Part 346 - LICENSES AND RADIATION SAFETY REQUIREMENTS FOR IRRADIATORS
Subpart C - DESIGN AND PERFORMANCE REQUIREMENTS OF IRRADIATORS
Section 346.250 - Shielding

Current through Register Vol. 48, No. 12, March 22, 2024

a) The radiation dose rate in areas that are normally occupied during operations of a panoramic irradiator may not exceed 0.02 millisievert (2 millirems) per hour at any location 30 centimeters or more from the wall of the room when the sources are exposed. The dose rate shall be averaged over an area not to exceed 100 square centimeters having no linear dimensions greater than 20 cm. Areas where the radiation dose rate exceeds 0.02 millisievert (2 millirems) per hour shall be locked, roped off or posted.

b) The radiation dose at 30 centimeters over the edge of the pool of a pool irradiator may not exceed 0.02 millisievert (2 millirems) per hour when the sources are in fully shielded position.

c) The radiation dose rate at 1 meter from the shield of a dry-source-storage panoramic irradiator when the source is shielded may not exceed 0.02 millisievert (2 millirems) per hour and at 5 centimeters from the shield may not exceed 0.2 millisievert (20 millirems) per hour.

Disclaimer: These regulations may not be the most recent version. Illinois may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.