Illinois Administrative Code
Title 32 - ENERGY
Part 340 - STANDARDS FOR PROTECTION AGAINST RADIATION
Subpart F - SURVEYS AND MONITORING
Section 340.540 - Calibration of Survey Instruments

Current through Register Vol. 48, No. 12, March 22, 2024

a) Unless specified in another Part, a licensee shall have each survey instrument used to show compliance with this Part calibrated before first use, annually and following a repair that affects the calibration. A licensee shall:

1) Calibrate all scales with readings up to 10 mSv (1000 mrem) per hour with a radiation source;

2) Calibrate two separated readings on each scale or decade that will be used to show compliance; and

3) Conspicuously note on the instrument the date of calibration.

b) A licensee may not use survey instruments if the difference between the indicated exposure rate and the calculated exposure rate is more than 20 percent.

Disclaimer: These regulations may not be the most recent version. Illinois may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.