Illinois Administrative Code
Title 32 - ENERGY
Part 335 - MEDICAL USE OF RADIOACTIVE MATERIAL
Subpart I - REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 335.8210 - Additional Technical Requirements for Mobile Remote Afterloader Units
Current through Register Vol. 48, No. 52, December 27, 2024
a) A licensee providing mobile remote afterloader service shall:
b) In addition to the periodic spot-checks required by Section 335.8170 of this Part, a licensee authorized to use mobile afterloaders for medical use shall perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of:
c) In addition to the requirements for checks in subsection (b) of this Section, a licensee shall ensure overall proper operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.
d) If the results of the checks required in subsections (b) and (c) of this Section indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace or check the malfunctioning system.
e) A licensee shall retain a record of each check for mobile remote afterloader units required by this Section for 5 years.
f) The record must include: