Illinois Administrative Code
Title 32 - ENERGY
Part 335 - MEDICAL USE OF RADIOACTIVE MATERIAL
Subpart I - REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 335.8170 - Periodic Spot-Checks for Remote Afterloader Units
Current through Register Vol. 48, No. 38, September 20, 2024
a) A licensee authorized to use a remote afterloader unit for medical use shall perform spot-checks of each remote afterloader facility and on each unit:
b) A licensee shall perform the measurements required by subsection (a) of this Section in accordance with written procedures established by the authorized medical physicist. That individual need not actually perform the spot-check measurements.
c) A licensee shall have the authorized medical physicist review the results of each spot-check within 15 days. The authorized medical physicist shall notify the licensee as soon as possible in writing of the results of each spot-check.
d) To satisfy the requirements of subsection (a) of this Section, spot-checks must, at a minimum, assure proper operation of:
e) If the results of the checks required in subsection (d) of this Section indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace or check the malfunctioning system.
f) A licensee shall retain a record of each spot-check for remote afterloader units required by this Section for 5 years.
g) The record must include, as applicable:
h) A licensee shall retain a copy of the procedures required by subsection (b) of this Section until the licensee no longer possesses the remote afterloader unit.