Illinois Administrative Code
Title 32 - ENERGY
Part 335 - MEDICAL USE OF RADIOACTIVE MATERIAL
Subpart I - REMOTE AFTERLOADER UNITS, TELETHERAPY UNITS AND GAMMA STEREOTACTIC RADIOSURGERY UNITS
Section 335.8150 - Full-Inspection Servicing for Teletherapy and Gamma Stereotactic Radiosurgery Units

Current through Register Vol. 48, No. 52, December 27, 2024

a) A licensee shall have each teletherapy unit and gamma stereotactic radiosurgery unit fully inspected and serviced during each source replacement to assure proper functioning of the source exposure mechanism and other safety components. The interval between each full-inspection servicing shall not exceed 5 years for each teletherapy unit and shall not exceed 7 years for each gamma stereotactic radiosurgery unit.

b) This inspection and servicing may only be performed by persons specifically licensed to do so by the Agency, the U.S. Nuclear Regulatory Commission, or an Agreement State.

c) A licensee shall maintain a record of the inspection and servicing for teletherapy and gamma stereotactic radiosurgery units required by this Section for the duration of use of the unit.

d) The record shall contain:

1) The inspector's radioactive materials license number;

2) The date of the inspection;

3) The manufacturer's name and model number and serial number of both the treatment unit and source;

4) A list of components inspected and serviced and the type of service; and

5) The signature of the inspector.

Disclaimer: These regulations may not be the most recent version. Illinois may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.