Section 335.7090 - Therapy-related Computer Systems for Manual Brachytherapy

Current through Register Vol. 48, No. 12, March 22, 2024

The licensee shall:

a) Perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:

1) The source-specific input parameters required by the dose calculation algorithm;

2) The accuracy of dose, dwell time, and treatment time calculations at representative points;

3) The accuracy of isodose plots and graphic displays; and

4) The accuracy of the software used to determine sealed source positions from radiographic images.

b) Maintain a record of acceptance testing and copies of the protocols used for acceptance testing in accordance with this Section for 5 years after discontinuation of use of the treatment planning system.

Disclaimer: These regulations may not be the most recent version. Illinois may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.