Illinois Administrative Code
Title 32 - ENERGY
Part 335 - MEDICAL USE OF RADIOACTIVE MATERIAL
Subpart C - GENERAL TECHNICAL REQUIREMENTS
Section 335.2010 - Possession, Use and Calibration of Instruments Used to Measure the Activity of Unsealed Radioactive Material

Current through Register Vol. 48, No. 12, March 22, 2024

a) For direct measurements performed in accordance with Section 335.2030, the licensee shall possess and use instrumentation to measure the activity of unsealed byproduct material before it is administered to each patient or human research subject.

b) A licensee shall calibrate the instrumentation required in subsection (a) in accordance with nationally recognized standards or the manufacturer's instructions.

c) A licensee shall maintain a record of instrument calibrations required by subsection (b) for 5 years. The records shall include the model and serial number of the instrument, the date of the calibration, the results of the calibration, the name of the individual who performed the calibration and a copy of the national standard or manufacturer's instructions used to perform the calibration.

Disclaimer: These regulations may not be the most recent version. Illinois may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.