Current through Rules and Regulations filed through December 18, 2024
(1)
Organic
Monitoring. Beginning on January 1, 1993, analysis of the contaminants
listed in Rule
391-3-5-.18(2)(b)1
-21 for the purpose of determining compliance with the maximum contaminant
level shall be conducted as follows:
(a)
Groundwater systems shall take a minimum of one sample at every entry point to
the distribution system which is representative of each well after treatment
(hereafter called a sampling point). Each sample must be taken at the same
sampling point unless conditions make another sampling point more
representative of each source, treatment plant, or within the distribution
system.
(b) Surface water systems
(or combined surface/ground) shall take a minimum of one sample at points in
the distribution system that are representative of each source or at each entry
point to the distribution system after treatment (hereafter called a sampling
point). Each sample must be taken at the same sampling point unless conditions
make another sampling point more representative of each source, treatment
plant, or within the distribution system.
(c) If the system draws water from more than
one source and the sources are combined before distribution, the system must
sample at an entry point to the distribution system during periods of normal
operating conditions (i.e., when water representative of all sources is being
used).
(d) Each community and
non-transient non-community water system shall take four consecutive quarterly
samples for each contaminant listed in Rule
391-3-5-.18(2)(b)2
-21 during each compliance period.
(e) If the initial monitoring for
contaminants listed in Rule
391-3-5-.18(2)(b)1
-8 and the monitoring for the contaminants listed in Rule
391-3-5-.18(2)(b)9
-21 as allowed in paragraph (1)(q) has been completed by December 31, 1992 and
the system did not detect any contaminant listed in Rule
391-3-5-.18(2)(b)1
-21, then each ground and surface water system shall take one sample
annually.
(f) After a minimum of
three years of annual sampling, the Division may allow groundwater systems with
no previous detection of any contaminant listed in Rule
391-3-5-.18(2)(b)
to take one sample during each compliance period.
(g) Each community and non-transient
groundwater system which does not detect a contaminant listed in Rule
391-3-5-.18(2)(b)1
-21 may apply to the Division for a waiver from the requirement of paragraph
(1)(e) and (1)(f) after completing the initial monitoring. (For the purposes of
paragraph (1), detection is defined as 0.0005 mg/L.) A waiver shall be
effective for no more than six years (two compliance periods). The Division may
also issue waivers to small systems for the initial round of monitoring for
1,2,4-trichlorobenzene.
(h) The
Division may grant a waiver after evaluating the factors in accordance with 40
CFR, Part 141.24(f)(8-9).
(i) Each
community and non-transient surface water system which does not detect a
contaminant listed in Rule
391-3-5-.18(2)(b)1
-21 may apply to the Division for a waiver from the requirements of paragraph
(1)(e) after completing the initial monitoring. Composite samples from a
maximum of five sampling points are allowed, provided that the detection limit
of the method used for analysis is less than one-fifth of the MCL. Systems
meeting this criteria must be determined by the Division to be non-vulnerable
based on a vulnerability assessment during each compliance period. Each system
receiving a waiver shall sample at the frequency specified by the Division (if
any).
(j) If a contaminant listed
in Rule 391-3-5-.18(2)(b)2
-21 is detected at a level exceeding 0.0005 mg/L in any sample, then:
1. The system must monitor quarterly at each
sampling point which resulted in a detection.
2. The Division may decrease the quarterly
monitoring requirements specified in paragraph (1)(j)(1); provided it has
determined that the system is reliably and consistently below the maximum
contaminant level. In no case shall the Division make this determination unless
a groundwater system takes a minimum of two quarterly samples and a surface
water system takes a minimum of four quarterly samples.
3. If the Division determines that the system
is reliably and consistently below the MCL, the Division may allow the system
to monitor annually. Systems which monitor annually must monitor during the
quarter(s) which previously yielded the highest analytical result.
4. Systems which have three consecutive
annual samples with no detection of a contaminant may apply to the Division for
a waiver as specified in paragraph (1)(g).
5. Groundwater systems which have detected
one or more of the following two-carbon organic compounds: trichloroethylene,
tetrachloroethylene, 1,2-dichloroethane, 1,1,1-trichloroethane,
cis-1,2-dichloroethylene, trans-1,2-dichloroethylene, or 1,1-dichloroethylene
shall monitor quarterly for vinyl chloride. A vinyl chloride sample shall be
taken at each sampling point at which one or more of the two-carbon organic
compounds was detected. If the results of the first analysis do not detect
vinyl chloride, the Division may reduce the quarterly monitoring frequency of
vinyl chloride monitoring to one sample during each compliance period. Surface
water systems are required to monitor for vinyl chloride as specified by the
Division.
(k) Systems
which violate the requirements of Rule
391-3-5-.18(2)(b)1
-21 as determined by paragraph (l)(n) must monitor quarterly. After a minimum
of four quarterly samples which show the system is in compliance as specified
in paragraph (l)(n), and the Division determines that the system is reliably
and consistently below the maximum contaminant level, the system may monitor at
the frequency and time specified in paragraph (l)(j)3.
(l) The Division may require a confirmation
sample for positive or negative results. If a confirmation sample is required
by the Division, the result must be averaged with the first sampling result and
the average is used for the compliance determination as specified by paragraph
(l)(n). The Division has the discretion to delete results of obvious sampling
errors from this calculation.
(m)
The Division may reduce the total number of samples a system must analyze by
allowing the use of compositing. Composite sampling and their analysis shall be
in accordance with 40 CFR, Part 141.24(f)(14).
(n) Compliance with Rule
391-3-5-.18(2)(b)1
-21 shall be determined based on the analytical results obtained at each
sampling point.
1. For systems which are
conducting monitoring at a frequency greater than annual, compliance is
determined by a running annual average of all samples taken at each sampling
point. If the annual average of any sampling point is greater than the MCL,
then the system is out of compliance. If the initial sample or a subsequent
sample would cause the annual average to be exceeded, then the system is out of
compliance immediately.
2. If
monitoring is conducted annually, or less frequently, the system is out of
compliance if the level of a contaminant at any sampling point is greater than
the MCL. If a confirmation sample is required by the Division, the
determination of compliance will be based on the average of two
samples.
3. If a public water
system has a distribution system separable from other parts of the distribution
system with no interconnections, the Division may allow the system to give
public notice to only that area served by that portion of the system which is
out of compliance.
4. If a system
fails to collect the required number of samples, compliance will be based on
the total number of samples collected.
(o) Analysis for the contaminants listed in
Rule 391-3-5-.18(2)(b)1
-21 shall be conducted in accordance with 40 CFR, Part 141.24(f)(17) . These
methods are contained in Methods for the Determination of Organic Compounds in
Drinking Water, EPA/600/4-88/039, December 1988 and are available from the
National Technical Information Service (NTIS) NTIS PB91-231480 and PB91-146027,
U.S. Department of Commerce, 5285 Port Royal Road, Springfield, Virginia
22161.
(p) Analysis under paragraph
(1) shall only be conducted by laboratories certified by the Division or
laboratories certified by EPA in accordance with conditions listed in 40 CFR,
Part 141.24(f)(17).
(q) The
Division may allow the use of monitoring data collected after January 1, 1998
required under section 1445 of the Public Health Service Act, as amended by the
Federal Safe Drinking Water Act, Public Law 93-523, for purposes of initial
monitoring compliance. If the data are generally consistent with the other
requirements in this rule, the Division may use these data (i.e., a single
sample rather than four quarterly samples) to satisfy the initial monitoring
requirement of paragraph (l)(d). Systems which use grandfathered samples and
did not detect any contaminant listed in Rule
391-3-5-.18(2)(b)2
-21 shall begin monitoring annually in accordance with paragraph
1(e).
(r) The Division may increase
required monitoring where necessary to detect variations within the
system.
(s) Each certified
laboratory must determine the method detection limit (MDL), as defined in 40
CFR, Part 136 appendix B, at which it is capable of detecting VOCs. The
acceptable MDL is 0.0005 mg/L. This concentration is the detection
concentration for purposes of paragraph (1).
(t) Each public water system shall monitor at
the time designated by the Division within each compliance period.
(2)
Initial Organic
Monitoring. For systems in operation before January 1, 1993, for
purposes of initial monitoring, analysis of the contaminants listed in Rule
391-3-5-.18(2)(b)1
-8 for purposes of determining compliance with the maximum contaminant levels
shall be conducted as follows:
(a)
Ground-water systems shall sample at points of entry to the distribution system
representative of each well after any application of treatment. Sampling must
be conducted at the same location(s) or more representative location(s) every
three months for one year except as provided in paragraph (2)(h).
(b) Surface water systems shall sample at
points in the distribution system representative of each source or at entry
points to the system after any application of treatment. Surface water systems
must sample each source every three months except as provided in paragraph
(2)(h). Sampling must be conducted at the same location or a more
representative location each quarter.
(c) If the system draws water from more than
one source and sources are combined before distribution, the system must sample
at an entry point to the distribution system during periods of normal operating
conditions.
(d) All community water
systems and non-transient, non-community water systems serving more than 10,000
people shall analyze all distribution or entry-point samples, as appropriate,
representing all source waters beginning no later than January 1, 1988. All
community water systems and non-transient, non-community water systems serving
from 3,300 to 10,000 people shall analyze all distribution or entry point
samples, as required in this paragraph (2), representing source waters no later
than January 1, 1989. All other community and non-transient, non-community
water systems shall analyze distribution or entry-point samples as required in
this paragraph (2), representing all source waters beginning no later than
January 1, 1991.
(e) The Division
may require confirmation samples for positive or negative results. If a
confirmation sample)s) is required by the Division, then the sample results(s)
should be averaged with the first sampling result and used for compliance
determination in accordance with paragraph (2)(i). The Division has the
discretion to delete results of obvious sampling errors from this
calculation.
(f) Analysis for vinyl
chloride is required only for ground water systems that have detected one or
more of the following two-carbon organic compounds: Trichloroethylene,
tetrachloroethylene, 1,2-dichloroethane, 1,1,1-trichloroethane,
cis-1,2-dichloroethylene, trans-1,2-dichloroethylene, or 1,1-dichloroethylene.
The analysis for vinyl chloride is required at each distribution or entry point
at which one or more of the two-carbon organic compounds were found. If the
first analysis does not detect vinyl chloride, the Division may reduce the
frequency of vinyl chloride monitoring to once every three years for that
sample location or other sample locations that are more representative of the
same source. Surface water systems may be required to analyze for vinyl
chloride at the discretion of the Division.
(g) The Division may allow compositing of up
to five samples from one or more public water systems.
(h) The Division may reduce the monitoring
frequency specified in paragraphs (2)(a) and (b) as explained in this
paragraph.
(i) Compliance with Rule
391-3-5-.18(2)(b)
shall be determined based on the results of running annual average of quarterly
sampling for each sampling location. If one location's average is greater than
the MCL, then the system shall be deemed to be out of compliance. If a public
water system has a distribution system separable from other parts of the
distribution system with no interconnections, only that part of the system that
exceeds any MCL as specified in Rule
391-3-5-.18(2)(b)
will be deemed out of compliance. The Division may reduce the public notice
requirement to that portion of the system which is out of compliance. If any
single sample result would cause the annual average to be exceeded, then the
system shall be deemed to be out of compliance immediately. For systems that
only take one sample per location because no VOCs were detected, compliance
shall be based on that one sample.
(j) Analysis under paragraph (2) shall only
be conducted by laboratories certified by the Division or have been certified
by the EPA.
(k) The Division may
allow the use of monitoring data collected after January 1, 1983, for purposes
of monitoring compliance. If the data is consistent with other requirements of
this rule, the Division may use that data to represent the initial monitoring
if the system is determined by the Division not to be vulnerable under the
requirements of paragraph (2). In addition, the result of EPA's Ground Water
Supply Survey may be used in a similar manner for systems supplied by a single
well.
(l) The Division may increase
required monitoring where necessary to detect variations within the
system.
(m) The Division may
determine compliance or initiate enforcement action based on analytical results
or other information compiled by their sanctioned representatives and
agencies.
(n) Each certified
laboratory must determine the method detection limit (MDL), as defined in 40
CFR, Part 136 appendix B, at which it is capable of detecting VOCs. The
acceptable MDL is 0.0005 mg/L. This concentration is the detection level for
purposes of paragraphs (2)(e), (f), and (g).
(3)
Ongoing Organic Monitoring.
Analysis of the contaminants listed in Rule
391-3-5-.18(2)(a)
for the purposes of determining compliance with the maximum contaminant level
shall be conducted as follows:
(a) Groundwater
systems shall take a minimum of one sample at every entry point to the
distribution system which is representative of each well after treatment
(hereafter called a sampling point). Each sample must be taken at the same
sampling point unless conditions make another sampling point more
representative of each source or treatment plant.
(b) Surface water systems shall take a
minimum of one sample at points in the distribution system that are
representative of each source or at each entry point to the distribution system
after treatment (hereafter called a sampling point). Each sample must be taken
at the same sampling point unless conditions make another sampling point more
representative of each source or treatment plant. [Note: For purposes of this
paragraph, surface water systems include systems with a combination of surface
and ground sources.]
(c) If the
system draws water from more than one source and the sources are combined
before distribution, the system must sample at an entry point to the
distribution system during periods of normal operating conditions (i.e., when
water representative of all sources is being used).
(d) Monitoring frequency:
1. Each community and non-transient
non-community water system shall take four consecutive quarterly samples for
each contaminant listed in Rule
391-3-5-.18(2)(a)
during each compliance period beginning with the compliance period starting
January 1, 1993.
2. Systems serving
more than 3,300 persons which do not detect a contaminant in the initial
compliance period, may reduce the sampling frequency to a minimum of two
quarterly samples in one year during each repeat compliance period.
3. Systems serving less than or equal to
3,300 persons which do not detect a contaminant in the initial compliance
period may reduce the sampling frequency to a minimum of one sample during each
repeat compliance period.
(e) Each community and non-transient water
system may apply to the Division for a waiver from the requirement of paragraph
(3)(d). A system must reapply for a waiver for each compliance
period.
(f) The Division may grant
a waiver after evaluating the factors in accordance with 40 CFR, Part
141.24(h)(6).
(g) If an organic
contaminant listed in Rule
391-3-5-.18(2)(a)
is detected (as defined by paragraph (3)(q)) in any sample, then:
1. Each system must monitor quarterly at each
sampling point which resulted in a detection.
2. The Division may decrease the quarterly
monitoring requirement specified in paragraph (3)(g)1 provided it has
determined that the system is reliably and consistently below the maximum
contaminant level. In no case shall the Division make this determination unless
a groundwater system takes a minimum of two quarterly samples and a surface
water system takes a minimum of four quarterly samples.
3. After the Division determines the system
is reliably and consistently below the maximum contaminant level the Division
may allow the system to monitor annually. Systems which monitor annually must
monitor during the quarter that previously yielded the highest analytical
result.
4. Systems which have three
(3) consecutive annual samples with no detection of a contaminant may apply to
the Division for a waiver as specified in paragraph (3)(f).
5. If monitoring results in detection of one
or more of certain related contaminants (aldicarb, aldicarb sulfone, aldicarb
sulfoxide and heptachlor, heptachlor epoxide), then subsequent monitoring shall
analyze for all related contaminants.
(h) Systems which violate the requirements of
Rule 391-3-5-.18(2)(a)
as determined by paragraph (3)(k) must monitor quarterly. After a minimum of
four quarterly samples show the system is in compliance and the Division
determines the system is reliably and consistently below the MCL, as specified
in paragraph (3)(k), the system shall monitor at the frequency specified in
paragraph (3)(g)3.
(i) The Division
may require a confirmation sample for positive or negative results. If a
confirmation sample is required by the Division, the result must be averaged
with the first sampling result and the average used for the compliance
determination as specified by paragraph (3)(k). The Division has the discretion
to delete results of obvious sampling errors from this calculation.
(j) The Division may reduce the total number
of samples a system must analyze by allowing the use of compositing. Composite
sampling and their analysis shall be in accordance with 40 CFR, Part
141.24(h)(10).
(k) Compliance with
Rule 391-3-5-.18(2)(a)
shall be determined based on the analytical results obtained at each sampling
point.
1. For systems which are conducting
monitoring at a frequency greater than annual, compliance is determined by a
running annual average of all samples taken at each sampling point. If the
annual average of any sampling point is greater than the MCL, then the system
is out of compliance. If the initial sample or a subsequent sample would cause
the annual average to be exceeded, then the system is out of compliance
immediately. Any samples below the detection limit shall be calculated as zero
for purposes of determining the annual average.
2. If monitoring is conducted annually, or
less frequently, the system is out of compliance if the level of a contaminant
at any sampling point is greater than the MCL. If a confirmation sample is
required by the Division, the determination of compliance will be based on the
average of two samples.
3. If a
public water system has a distribution system separable from other parts of the
distribution system with no interconnections, the Division may allow the system
to give public notice to only that portion of the system which is out of
compliance.
4. If a system fails to
collect the required number of samples, compliance will be based on the total
number of samples collected.
(l) Analysis for the contaminants listed in
Rule 391-3-5-.18(2)(a)
shall be conducted in accordance with 40 CFR, Part 141.24(h) (12-13). These
methods are contained in "Methods for the Determination of Organic Compounds in
Drinking Water", ORD Publications, CERI, EPA/600/4-88/039, December
1988.
(m) If monitoring data
collected after January 1, 1990, are generally consistent with the requirements
of Rule 391-3-5-.22(3),
then the Division may allow systems to use that data to satisfy the monitoring
requirement for the initial compliance period beginning January 1,
1993.
(n) The Division may increase
the required monitoring frequency, where necessary, to detect variations within
the system (e.g., fluctuations in concentration due to seasonal use, changes in
water source).
(o) The Division has
the authority to determine compliance or initiate enforcement action based upon
analytical results and other information compiled by their sanctioned
representatives and agencies.
(p)
Each public water system shall monitor at the time designated by the Division
within each compliance period.
(q)
Detection limits for contaminants used in paragraph (3) shall be in accordance
with 40 CFR, Part 141.24(h)(18).
(r) Analysis under paragraph (3) shall
conform to paragraph (1) of Rule
391-3-5-.29.
(s) The best technology, treatment technique,
or other means available for achieving compliance with the maximum contaminant
level for organic contaminants in Rule
391-3-5-.18(2)(a) and
(2)(b) shall be in accordance with 40 CFR,
Part 141.61(b).
O.C.G.A.
§
12-5-170
et
seq.