Rule 276-5-.03 - Calibration

Universal Citation: GA Rules and Regs r 276-5-.03

Current through Rules and Regulations filed through March 20, 2024

Licensees shall use calibrated equipment in the delivery of professional services and shall maintain calibration reports for six (6) years. Equipment shall be calibrated annually, or more frequently if in accordance with the manufacturer's specifications, and shall have calibrator's certification that post-calibration readings meet or exceed American National Standards Institute ("ANSI") standards. Calibration reports shall contain as a minimum the following:

(a) Audiometer's identification including make, model, and serial number;

(b) Calibrator's identification including company name and location and name of technician responsible for work;

(c) Post calibration readings for air and bone conduction, speech, signal leakage, attenuation linearity, and masking;

(d) Calibrator's certification that post calibration readings meet or exceed ANSI standards;

(e) Date of calibration.

O.C.G.A. Sec. 43-20-6.

Disclaimer: These regulations may not be the most recent version. Georgia may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.