Compilation of Rules and Regulations of the State of Georgia
Department 276 - GEORGIA STATE BOARD OF HEARING AID DEALERS AND DISPENSERS
Chapter 276-5 - PROCEDURES, EQUIPMENT, CALIBRATIONS AND RECORDS
Rule 276-5-.03 - Calibration
Current through Rules and Regulations filed through March 20, 2024
Licensees shall use calibrated equipment in the delivery of professional services and shall maintain calibration reports for six (6) years. Equipment shall be calibrated annually, or more frequently if in accordance with the manufacturer's specifications, and shall have calibrator's certification that post-calibration readings meet or exceed American National Standards Institute ("ANSI") standards. Calibration reports shall contain as a minimum the following:
(a) Audiometer's identification including make, model, and serial number;
(b) Calibrator's identification including company name and location and name of technician responsible for work;
(c) Post calibration readings for air and bone conduction, speech, signal leakage, attenuation linearity, and masking;
(d) Calibrator's certification that post calibration readings meet or exceed ANSI standards;
(e) Date of calibration.
O.C.G.A. Sec. 43-20-6.