Instruments used in hematological examination of specimens shall be recalibrated, retested or reinspected, as appropriate, each day of use. Each procedure shall be recalibrated or rechecked each shift of use with standards or controls covering the entire range of expected values, unless required more frequently by the manufacturer or federal laboratory regulations. Tests such as the hematocrit and one-stage prothrombin time test shall be run in duplicate except as specified in published guidelines. Standard deviation, coefficient of variation, or other statistical estimates of precision shall be determined by the laboratory. All control materials used to satisfy the control requirement must have documented established limits.

O.C.G.A. § 31-22-1et seq.

Disclaimer: These regulations may not be the most recent version. Georgia may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.