Florida Administrative Code
64 - DEPARTMENT OF HEALTH
64E - Division of Environmental Health
Chapter 64E-5 - CONTROL OF RADIATION HAZARDS
Section 64E-5.6331 - Calibration Measurements of Manual Brachytherapy Sources
Current through Reg. 50, No. 187; September 24, 2024
(1) Before the first medical use of a brachytherapy source, the licensee shall, using published protocols currently accepted by nationally recognized bodies, determine the following:
(2) Instead of a licensee making its own measurements as required in subsection 64E-5.6331(1), F.A.C., the licensee may use measurements provided by the source manufacturer or by a calibration laboratory accredited by the American Association of Physicists in Medicine (AAPM) that are made in accordance with subsection 64E-5.6331(1), F.A.C.
(3) A licensee shall mathematically correct the outputs or activities determined in subsection 64E-5.6331(1), F.A.C., for physical decay at intervals consistent with 1 percent physical decay.
(4) For each brachytherapy source the licensee shall retain the following records for three years after the last use of the source:
Rulemaking Authority 404.022, 404.051, 404.061, 404.071, 404.081, 404.141 FS. Law Implemented 404.022, 404.051(1), (4), (5), (6), (8), (9), (10), (11), 404.061(2), (3), 404.071(1), 404.081, 404.141 FS.
New 2-11-10.