Connecticut Administrative Code
Title 19 - Public Health and Safety
25d - X-Ray Devices used for Diagnosis and Therapy
Section 19-25d-6 - Special requirements for mobile diagnostic radiographic equipment
Universal Citation: CT Reg of State Agencies 19-25d-6
Current through September 9, 2024
(a) Equipment
(1) All requirements of section
19-25d-5
apply except subdivision (a) (5).
(2) The exposure control switch shall be of
the dead-man type and shall be so arranged that the operator can stand at least
six feet from the patient and well away from the useful beam.
(b) Structural shielding When a mobile unit is used routinely in one location, it shall be considered a fixed installation subject to the shielding requirements specified in sections 19-25d-3(c) and 19-25d-5(b).
(c) Operating procedures
(1) All provisions of subsection
19-25d-5(c)
apply except subdivision (2).
(2)
The target-to-skin distance shall be not less than twelve inches.
(3) Personnel monitoring shall be required
for all individuals operating mobile x-ray equipment.
Disclaimer: These regulations may not be the most recent version. Connecticut may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.