Connecticut Administrative Code
Title 19 - Public Health and Safety
25d - X-Ray Devices used for Diagnosis and Therapy
Section 19-25d-5 - Radiographic installation other than dental and veterinary medicine
Universal Citation: CT Reg of State Agencies 19-25d-5
Current through September 9, 2024
(a) Equipment
(1) The tube housing shall be of diagnostic
type.
(2) Diaphragms or cones
capable of restricting the beam to the area of clinical interest shall be
provided for collimating the useful beam and shall provide the same degree of
protection as is required of the housing.
(3)
(A)
Except when contraindicated for a particular medical purpose, for equipment
operating at seventy kvp and below, the total filtration permanently in the
useful beam shall be equivalent to at least one and one-half mm of aluminum.
This requirement may be assumed to have been met if the half-value layer is not
less than one and one-half mm aluminum at normal operating voltages.
(B) Except when contraindicated for a
particular medical purpose, for equipment capable of operating above seventy
kvp, the total filtration permanently in the useful beam shall be equivalent to
at least two and one-half value layer is not less than two and one-half mm
aluminum at normal operating voltages.
(4) A device shall be provided to terminate
the exposure after a preset time or exposure.
(5) A dead-man type of exposure switch shall
be so arranged that it cannot be conveniently operated outside a shielded area.
Exposure switches for "spot film" devices used in conjunction with fluoroscopic
tables are excepted from this shielding requirement.
(b) Structural shielding
(1) All wall, floor and ceiling areas exposed
to the useful beam shall have primary barriers. Primary barriers in walls shall
extend to a minimum height of eighty-four inches above the floor.
(2) Secondary barriers shall be provided in
all wall, floor and ceiling areas not having primary barriers or where the
primary barrier requirements are lower than the secondary barrier
requirements.
(3) The operator's
station at the control shall be behind a protective barrier, either in a
separate room, in a protected booth, or behind a shield which will intercept
the useful beam and any radiation which has been scattered only once.
(4) A window of lead-equivalent glass equal
to that required by the adjacent barrier or a mirror system shall be provided
large enough and so placed that the operator can see the patient without having
to leave the protected area during exposure.
(c) Operating procedures
(1) No individual occupationally exposed to
radiation shall be permitted to hold patients during exposures except during
emergencies, nor shall any individual be regularly used for this
service.
(2) Only individuals
required for the radiographic procedure shall be in the radiographic room
during exposure; and, except for the patient, no unprotected parts of their
bodies shall be in the useful beam.
(3) The useful beam shall be restricted to an
area of clinical interest.
Disclaimer: These regulations may not be the most recent version. Connecticut may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.