Code of Colorado Regulations
1000 - Department of Public Health and Environment
1007 - Hazardous Materials and Waste Management Division
6 CCR 1007-1 - RULES AND REGULATIONS PERTAINING TO THE ADMINISTRATION AND ENFORCEMENT OF THE PESTICIDE APPLICATORS' ACT
Part 24 - RADIATION CONTROL - PARTICLE ACCELERATORS AND THERAPEUTIC RADIATION MACHINES IN THE HEALING ARTS
SPECIFIC REQUIREMENTS
Section 6 CCR 1007-1-24.9 - Quality Assurance For Radiation Therapy Simulation Systems
Universal Citation: 1000 CO Code Regs 6 CCR 1007-1-24.9
Current through Register Vol. 48, No. 6, March 25, 2025
24.9.1 Quality assurance for a simulator shall include acceptance testing and periodic verification of system performance; and
24.9.2 Be performed (unless the Registered Medical Physicist determines and documents that a particular recommendation is not warranted for the clinical task for which the equipment will be used) in accordance with:
24.9.2.1 AAPM Report 46, for a radiographic/fluoroscopic simulator; or
24.9.2.2 AAPM Report 83 for a computed tomography based simulator.
Disclaimer: These regulations may not be the most recent version. Colorado may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.